This paper was converted on www.awesomepapers.org from LaTeX by an anonymous user.
Want to know more? Visit the Converter page.

Non-invasive quantitative imaging of selective microstructure-sizes with magnetic resonance

Milena Capiglioni Centro Atómico Bariloche, CONICET, CNEA, S. C. de Bariloche, 8400, Argentina Instituto Balseiro, CNEA, Universidad Nacional de Cuyo, S. C. de Bariloche, 8400, Argentina Support Center for Advanced Neuroimaging (SCAN), Institute of Diagnostic and Interventional Neuroradiology, University of Bern, Bern, 3010 , Switzerland    Analia Zwick [email protected] Centro Atómico Bariloche, CONICET, CNEA, S. C. de Bariloche, 8400, Argentina Departamento de Física Médica, Instituto de Nanociencia y Nanotecnologia, CNEA, CONICET, S. C. de Bariloche, 8400, Argentina    Pablo Jiménez Centro Atómico Bariloche, CONICET, CNEA, S. C. de Bariloche, 8400, Argentina Instituto Balseiro, CNEA, Universidad Nacional de Cuyo, S. C. de Bariloche, 8400, Argentina    Gonzalo A. Álvarez [email protected] Centro Atómico Bariloche, CONICET, CNEA, S. C. de Bariloche, 8400, Argentina Instituto Balseiro, CNEA, Universidad Nacional de Cuyo, S. C. de Bariloche, 8400, Argentina Departamento de Física Médica, Instituto de Nanociencia y Nanotecnologia, CNEA, CONICET, S. C. de Bariloche, 8400, Argentina
Abstract

Extracting reliable and quantitative microstructure information of living tissue by non-invasive imaging is an outstanding challenge for understanding disease mechanisms and allowing early stage diagnosis of pathologies. Magnetic Resonance Imaging is the favorite technique to pursue this goal, but still provides resolution of sizes much larger than the relevant microstructure details on in-vivo studies. Monitoring molecular diffusion within tissues, is a promising mechanism to overcome the resolution limits. However, obtaining detailed microstructure information requires the acquisition of tens of images imposing long measurement times and results to be impractical for in-vivo studies. As a step towards solving this outstanding problem, we here report on a method that only requires two measurements and its proof-of-principle experiments to produce images of selective microstructure sizes by suitable dynamical control of nuclear spins with magnetic field gradients. We design microstructure-size filters with spin-echo sequences that exploit magnetization “decay-shifts” rather than the commonly used decay-rates. The outcomes of this approach are quantitative images that can be performed with current technologies, and advance towards unravelling a wealth of diagnostic information based on microstructure parameters that define the composition of biological tissues.

The development of nanosized sensors with novel quantum technologies is aiming at nanoscale imaging of biological tissues for unveiling the biophysics of pathologies at such relevant scales (Staudacher et al., 2015; Wang et al., 2019; Barry et al., 2016; Glenn et al., 2015). Such imaging proposals are still based on invasive techniques. By contrast, Magnetic Resonance Imaging (MRI) has proven to be an excellent tool for acquiring non-invasive images, being applied on a daily basis for clinical diagnosis. However, the weak sensitivity for detecting the nuclear spins inherent to the biological tissues, typically limits the spatial resolution of in-vivo MRI to hundred of micrometers in pre-clinical scanners and to millimeters in clinical systems. This limitation imposes a challenge for existing methods to early detect diseases that produce changes at the cellular level (Padhani et al., 2009; Drago et al., 2011; White et al., 2013, 2014; Enzinger et al., 2015). Detecting these kind of pathologies in a development stage based on quantitative imaging of tissue microstructure parameters, will allow MRI to advance towards a new early diagnostic paradigm (Le Bihan, 2003; White et al., 2013, 2014; Xu et al., 2014; Enzinger et al., 2015; Grussu et al., 2017; Alexander et al., 2019).

Diffusion Weighted MRI (DWI) is a promising tool to probe microstructure information based on monitoring the dephasing of the nuclear spin precession due to Brownian molecular motion (Le Bihan, 2003; Grebenkov, 2007; Callaghan, 2011). The diffusion dynamics of molecules depends on tissue properties such as cell sizes, density, and other morphological features. A strong magnetic field gradient is applied to sense the microscopic motion of spins so that the the precession frequency depends on their instantaneous position. In addition, modulating the gradient strength as a function of time allows to probe the time dependent diffusion process of the molecules within tissues (Stepisnik, 1993; Callaghan, 2011; Álvarez et al., 2013; Shemesh et al., 2013). These dynamical control techniques are based on the Hahn spin-echo concept (Hahn, 1950) and its generalization to multiple echoes (Carr and Purcell, 1954). Within DWI they are called modulated gradient spin-echo (MGSE) sequences, where the phase accumulated by the spin’s precession is refocused at given times allowing to infer microscopic parameters in tissues and porous media (Stepisnik, 1993; Shemesh et al., 2013; Drobnjak et al., 2016; Nilsson et al., 2017; Novikov et al., 2019). However, obtaining detailed microstructure information is still challenging on in-vivo studies, as it requires tens of images demanding about an hour of acquisition time (Assaf et al., 2008; Alexander et al., 2010; Xu et al., 2014; Shemesh et al., 2015; Alexander et al., 2019; Novikov et al., 2019).

The decay of the nuclear spin signal under MGSE sequences is typically characterized by a decay-rate (Grebenkov, 2007; Callaghan, 2011). Here, we report that these decaying signals manifest a “decay-shift” that can be exploited to selectively probe microstructure-sizes. We develop size-filters based on the Non-uniform Oscillating Gradient Spin-Echo (NOGSE) concept that contrast the signal generated by two spin-echo sequences (Álvarez et al., 2013; Shemesh et al., 2013). The method probes different diffusion time scales while factoring out other relaxation mechanisms induced by gradient-modulation imperfections and T2T_{2} effects. We exploit the NOGSE modulations to selectively probe the spin-echo decay-shift for producing quantitative images based on contrast intensities that reflect the probability of finding a specific microstructure-size. The present approach requires only two measurements, therefore significantly reducing the acquisition time compared to state-of-the-art methods that typically require tens of measurements to obtain quantitative information on microstructure-sizes based on fitting parameters (Assaf et al., 2008; Alexander et al., 2010; Ong and Wehrli, 2010; Xu et al., 2014; Shemesh et al., 2015; Alexander et al., 2019; Novikov et al., 2019). We analytically and experimentally demonstrate that these microstructure-size filters can be implemented with current technologies presenting a novel mechanism for quantitative and precision imaging diagnostic tools.

MRI of molecular-diffusion. The nuclear spins S=12S\!=\!\frac{1}{2} in molecules intrinsic to biological tissues, mainly from water’s protons, are typically observed on in-vivo MRI. They interact with an external uniform magnetic-field B0B_{0}z^\hat{z} and a magnetic field gradient Gr^G\hat{r} applied along the direction r^\hat{r} for spatial encoding. In a frame rotating at the resonance frequency γB0\gamma B_{0}, the precession frequency ω(t)=γGr(t)\omega(t)=\gamma Gr(t) fluctuates reflecting the random motion of the molecular diffusion process (Grebenkov, 2007; Callaghan, 2011). Here, r(t)r(t) is the instantaneous position of the nuclear spin along the gradient direction and γ\gamma is the gyromagnetic ratio of the nucleus. The effective gradient strength may vary along time as G(t)G(t), either by applying π\pi-pulses if the gradient is constant or directly by modulating the gradient’s sign and amplitude. The spins dephasing induced by the diffusion process is refocused by these control modulations forming the so-called spin-echoes (Hahn, 1950; Carr and Purcell, 1954) (Fig. 1a,b). At the evolution time TET_{E} of the control sequence, the spin-echo magnetization decays depending on how nuclear spins were scrambled by the diffusion process. The resulting magnetization at TET_{E} is then spatially encoded with an MRI acquisition sequence.

The spins acquire a random phase ϕ(TE)\phi(T_{E}) that typically follows a Gaussian distribution (Stepisnik, 1999). The magnetization in a voxel of the image becomes

M(TE)=e12ϕ2(TE)M(0).M(T_{E})=e^{-\frac{1}{2}\left\langle\phi^{2}(T_{E})\right\rangle}M(0). (1)

The quadratic phase is averaged over the spin ensemble as (Stepisnik, 1993; Grebenkov, 2007; Callaghan, 2011)

ϕ2(TE)=γ20TE𝑑t0TE𝑑tG(t)G(t)Δr(tt)Δr(0),\left\langle\phi^{2}(T_{E})\right\rangle\!=\!\gamma^{2}\!\int_{0}^{T_{E}}\!\!\negthinspace\!dt\!\int_{0}^{T_{E}}\!\!\!dt^{\prime}G(t)G(t^{\prime})\left\langle\Delta r(t-t^{\prime})\Delta r(0)\right\rangle, (2)

which is expressed in terms of the control applied to the system by G(t)G(t) and the molecular displacement autocorrelation function Δr(0)Δr(t)=D0τce|t|/τc\left\langle\Delta r(0)\Delta r(t)\right\rangle=D_{0}\tau_{c}e^{-|t|/\tau_{c}} (Stepisnik, 1993; Álvarez et al., 2013; Shemesh et al., 2013). Here Δr(t)=r(t)r(t)\Delta r(t)=r(t)-\left\langle r(t)\right\rangle is the instantaneous displacement of the spin position from its mean value, D0D_{0} is the free diffusion coefficient, and τc\tau_{c} is the correlation time of the diffusion process. The restriction length lcl_{c} of the microstructure compartment in which molecular diffusion is taking place is then determined by the Einstein-Diffusion equation lc2=D0τcl_{c}^{2}=D_{0}\tau_{c} (Callaghan, 2011). Then, by monitoring the spin-echo decay by applying suitable control sequences, one can infer microstructure-sizes (Ong and Wehrli, 2010; Álvarez et al., 2013; Shemesh et al., 2013; Drobnjak et al., 2016; Nilsson et al., 2017; Novikov et al., 2019).

Refer to caption
Figure 1: Spin-echo decay-shift as a paradigm for probing microstructure sizes. (a,b) Magnetic resonance spin-echo sequences for DWI. An initial π2\frac{\pi}{2}-excitation pulse is followed by a constant magnetic field gradient GG. During the evolution time TET_{E}, equidistant π\pi rf-pulses modulate the effective gradient G(t)G(t), switching its sign for refocusing the spins dephasing induced by the diffusion process to form the spin-echoes. (a) Hahn sequence with a refocusing period tHt_{H} (N=1N\!=\!1). (b) CPMG sequence with N>1N\!>\!1 refocusing periods of duration tCt_{C}. After TET_{E} the remaining signal is measured, possibly using MRI acquisition encoding. (c) Spin-echo decay for the Hahn (N=1N\!=\!1, orange line) and the CPMG (N=8N\!=\!8, green line) gradient modulation as a function of the normalized evolution time TE/τcT_{E}/\tau_{c}, where τc\tau_{c} is the correlation time of the restricted diffusion process. For both sequences, when the refocusing period is TE/NτcT_{E}/N\gg\tau_{c}, the decaying signal exp(γ2G2D0τc2TE)\propto\exp\left(-\gamma^{2}G^{2}D_{0}\tau_{c}^{2}\,T_{E}\right) (black solid-lines) has a constant decay-rate independent of NN. However, the curves have a decay-shift exp(γ2G2D0τc3(1+2N))\propto\exp\left(\gamma^{2}G^{2}D_{0}\tau_{c}^{3}(1+2N)\right) independent of time but depending on NN (marked with arrows) with respect to the dashed-line that gives exp(γ2G2D0τc2TE)\exp\left(-\gamma^{2}G^{2}D_{0}\tau_{c}^{2}\,T_{E}\right). The contrast ΔM\Delta M between the CPMG and Hahn decay is highlighted in the plot with double-arrow. The inset shows a scheme for molecules undergoing restricted Brownian motion. The diffusion restriction length lc=D0τcl_{c}=\sqrt{D_{0}\tau_{c}} is related to the compartment-size. (d) A CPMG sequence of N1N\!-\!1 equidistant refocusing periods tCt_{C} is concatenated with a Hahn sequence of refocusing period tHt_{H} (Álvarez et al., 2013). The effective modulated gradient G(t)G(t) is shown at the bottom, and it composes the Non-uniform Oscillating-Gradient Spin-Echo (NOGSE) sequence that can be generated by directly modulating the gradient strength (Shemesh et al., 2013). (e) NOGSE contrast (NOGSEc) ΔM\Delta M as a function of the normalized evolution time TEτc\frac{T_{E}}{\tau_{c}} for Lc6=γ2G2D0τc3=0.01L_{c}^{6}\!=\!\gamma^{2}G^{2}D_{0}\tau_{c}^{3}=0.01. This value is representative of white-matter tissue considering D0=0.7μm2/msD_{0}\!=\!0.7\,\mu\mathrm{m^{2}/ms} and τc=1.5ms\tau_{c}\!=\!1.5\,\mathrm{ms} with G=240mT/mG\!=\!240\,\mathrm{mT/m} and N=8N\!=\!8.

Spin-echo decay-shift as a probing-microstructure paradigm. The spin-echo magnetization decay is usually characterized by its decay-rate (Grebenkov, 2007; Callaghan, 2011). Under the control sequences discussed in Fig. 1a,b, the decay-rate is typically reduced as the number of refocusing periods NN increases (Carr and Purcell, 1954). This effect is shown in Fig. 1a-c, where the Hahn spin-echo (N=1)(N\!=\!1) decay is compared with the signal after a CPMG sequence (N>1N\!>\!1) with multiple echoes (Carr and Purcell, 1954). However, in the restricted diffusion regime when the refocusing periods are longer than the correlation time τc\tau_{c}, the spin-echo decays as M(TE)/M(0)eγ2G2D0τc2(TE(1+2N)τc)M(T_{E})/M(0)\approx e^{-\gamma^{2}G^{2}D_{0}\tau_{c}^{2}(T_{E}-(1+2N)\tau_{c})} (see Methods). There, the spins have been fully scrambled within the compartment, and the dephasing cannot be refocused leading to a decay-rate γ2G2D0τc2\gamma^{2}G^{2}D_{0}\tau_{c}^{2} independent of NN. Yet, the spin-echo retains information of the transition from the free to the restricted diffusion regime. This information is manifested as a “decay-shift” γ2G2D0τc3(1+2N)\gamma^{2}G^{2}D_{0}\tau_{c}^{3}(1+2N) independent of the evolution time on the spin-echo decay signal, as shown in Fig. 1c. We here demonstrate that this shift can be exploited to selectively probe microstructure sizes as it is lc6\propto l_{c}^{6}.

As this decay-shift depends on NN, it can be selectively probed by concatenating a Hahn with a CPMG gradient modulation and changing the ratio between the relative refocusing periods of each component of the sequence as shown in Fig. 1d. This control sequence produces an effective modulated gradient G(t)G(t), that conforms the Non-uniform Oscillating-Gradient Spin-Echo (NOGSE) (Álvarez et al., 2013; Shemesh et al., 2013). This sequence also factorizes out other relaxation mechanisms allowing to probe selectively the diffusion induced decay.

We define the NOGSE contrast (NOGSEc) as the amplitude ΔM\Delta M given by the difference between the CPMG and the Hahn signal (Fig. 1c). It is obtained by evaluating NOGSE at the refocusing periods tC=tHt_{C}=t_{H} and then at tC0t_{C}\rightarrow 0 using the definitions shown in Fig. 1d. Then both measurements are subtracted (see Methods). Within the restricted diffusion regime, this contrast amplitude is

ΔMeγ2G2D0τc3(TEτc3)(eγ2G2D0τc32(N1)1),\Delta M\approx e^{-\gamma^{2}G^{2}D_{0}\tau_{c}^{3}(\frac{T_{E}}{\tau_{c}}-3)}\!\left(e^{\gamma^{2}G^{2}D_{0}\tau_{c}^{3}2(N-1)}-1\right), (3)

which is very sensitive to the restricted diffusion length lcl_{c} as it has a parametric dependence lc6τc3l_{c}^{6}\propto\tau_{c}^{3} provided by the spin-echo decay-shift (Álvarez et al., 2013; Shemesh et al., 2013).

NOGSE as a selective microstructure-size filter. NOGSEc ΔM\Delta M has a maximum as a function of the normalized echo-time TE/τcT_{E}/\tau_{c} as shown in Fig. 1e. We exploit this maximum contrast to enhance the relative contribution to the signal from specific restriction lengths lcl_{c} from a size-distribution.

In order to perform a general analysis, Eq. (3) can be expressed in terms of dimensionless lengths LD=lD/lG,Lc=lc/lGL_{D}=l_{D}/l_{G},L_{c}=l_{c}/l_{G} (see Methods). Here, lc=D0τc,lD=D0TE,lG=D0/γG3l_{c}=\sqrt{D_{0}\tau_{c}},\>l_{D}=\sqrt{D_{0}T_{E}},\>l_{G}=\sqrt[3]{D_{0}/\gamma G} are the restriction length, the diffusion length that the spin can diffuse freely during TET_{E} and the dephasing diffusion length that provides a phase shift of 2π2\pi, respectively (Callaghan, 2011) (Fig. 1c). Then, ΔM\Delta M as a function of LcL_{c} can be approximated by a Gaussian function when LD/Lc1L_{D}/L_{c}\gg 1, Lc61L_{c}^{6}\ll 1 and LD1L_{D}\gg 1 (see Methods),

ΔM2(N1)e3/2(Lcf)6exp[12(LcLcfLcf)2].\Delta M\approx 2(N-1)e^{-3/2}\left(L_{c}^{f}\right)^{6}\exp\left[-12\left(\frac{L_{c}-L_{c}^{f}}{L_{c}^{f}}\right)^{2}\right]. (4)

NOGSEc therefore acts as a microstructure-size “bandpass-filter” with LcfL_{c}^{f} as the filter-center size (see Fig. 2a)

Lcf(3/2)14LD12.L_{c}^{f}\approx\left(3/2\right)^{\frac{1}{4}}L_{D}^{-\frac{1}{2}}. (5)

The filter-band selectivity is defined by the ratio between the full-width-at-half-maximum (FWHM) and LcfL_{c}^{f}, FWHM/Lcfln230.5FWHM/L_{c}^{f}\approx\sqrt{\frac{\ln 2}{3}}\approx 0.5.

The maximum of ΔM\Delta M at the size LcfL_{c}^{f} can be tuned to highlight a given restriction length lcl_{c} based on choosing properly the sequence control parameters, i.e. the gradient strength GG and the evolution time TET_{E} (see Fig. 2a).

Refer to caption
Figure 2: Selective microstructure-size filter based on NOGSE contrast. (a) NOGSEc amplitude ΔM\Delta M as a function of the renormalized restriction length LcL_{c} for different values of the diffusion length LDL_{D} (LD2=15,20L_{D}^{2}\!=\!15,20 with N=8N\!=\!8 refocusing periods). ΔM\Delta M shows a Gaussian filter functional dependence for LD1L_{D}\gg 1 with a maximum at LcfLD12L_{c}^{f}\propto L_{D}^{-\frac{1}{2}}. The filter full-width-half-maximun FWHMLcf2FWHM\approx\frac{L_{c}^{f}}{2}. (b) The amplitude of ΔM\Delta M at LcfL_{c}^{f} increases linearly with the refocusing periods NN as long as LD2Lc2N1\frac{L_{D}^{2}}{L_{c}^{2}N}\gg 1. Here LD2=22L_{D}^{2}=22. (c) Attenuation effects of the filter amplitude due to transversal T2T_{2}-relaxation. Different dephasing diffusion lengths L22=10,50,L_{2}^{2}\!=\!10,50,\infty associated to T2T_{2} are shown for LD2=20L_{D}^{2}\!=\!20. ΔM\Delta M decreases with decreasing L2L_{2}. (d,e) ΔM\Delta M as a function of the restriction length lcl_{c} and the gradient strength GG, considering N=8N\!=\!8. As LD2=25L_{D}^{2}\!=\!25 remains constant by properly varying TET_{E}, the maximum ΔM\Delta M is also constant. The considered dynamic range for the gradient strength GG is achievable with current technologies. The horizontal dashed lines in (e) correspond to the specific cases plotted in (d). LcfL_{c}^{f} increases with decreasing GG. D0=0.7μm2/msD_{0}\!=\!0.7\,\mu\mathrm{m^{2}/ms} is considered.

The filtered size decreases with increasing the control parameter LDL_{D} according to Eq. (5). At the same time, increasing LDL_{D} decreases the contrast amplitude which is (Lcf)61/LD3\propto\left(L_{c}^{f}\right)^{6}\propto 1/L_{D}^{3}. Therefore, the minimum size that can be filtered in practice is limited by the Signal-to-Noise Ratio (SNR) and the maximum achievable gradient strength as LDG3L_{D}\propto\sqrt[3]{G}. This decrease of ΔM\Delta M can be compensated linearly with increasing the number of refocusing periods NN, as long as LD2Lc2N=TEτcN1\frac{L_{D}^{2}}{L_{c}^{2}N}=\frac{T_{E}}{\tau_{c}N}\gg 1 to reach the restricted regime (see Eq. (4) and Fig. 2b).

A practical limitation is also the unavoidable transversal T2T_{2}-relaxation due to the intrinsic dephasing of the nuclear spins. NOGSEc decays then as

ΔMT2=ΔMeLD2/L22,\Delta M_{T_{2}}=\Delta M\,e^{-L_{D}^{2}/L_{2}^{2}}, (6)

where we have defined L2=l2lGL_{2}=\frac{l_{2}}{l_{G}} as the T2T_{2} diffusion dimensionless length with l2=(D0T2)1/2l_{2}=(D_{0}T_{2})^{1/2}. The T2T_{2}-relaxation effect is showed in Fig. 2c for different values of L22L_{2}^{2}. Remarkably, the filter shape and center remains the same as the case of T2T_{2}\rightarrow\infty.

Selective size-filtering in typical microstructure-size distributions. Our results show that one can produce quantitative images based on a signal contrast generated by specific microstructure-sizes from a size-distribution. This method avoids extracting the microstructure-size by fitting a curve, which is time consuming as it typically requires several measurements (Assaf et al., 2008; Alexander et al., 2010; Ong and Wehrli, 2010; Xu et al., 2014; Shemesh et al., 2015; Novikov et al., 2019). The filter amplitude ΔM\Delta M in Eq. (4) remains constant by varying the gradient strength and the evolution time keeping fixed the parameter LDL_{D}, as shown in Fig. 2d-e. We exploit this property for a proof-of-principle evaluation of the performance of the NOGSEc filter applied to microstructural size-distribution inherent to heterogeneous biological tissues as shown in Fig. 3. Typical size-distributions P(lc)P(l_{c}) are log-normal and bi-modal functions (Assaf et al., 2008; Pajevic and Basser, 2013; White et al., 2013; Liewald et al., 2014; Shemesh et al., 2015) as shown in Figs. 3a-b. NOGSEc for a size-distribution is given by

ΔM=lcP(lc)ΔM(lc)𝑑lc,\Delta M=\int_{l_{c}}P(l_{c})\Delta M(l_{c})dl_{c}, (7)

where ΔM(lc)\Delta M(l_{c}) is the contribution for a given restriction length lcl_{c}. Figures 3c-d show ΔM\Delta M as a function of lcf=LcflGl_{c}^{f}=L_{c}^{f}l_{G}, where GG and TET_{E} are changed simultaneously keeping LD2L_{D}^{2} constant. The center of the filter at lcfl_{c}^{f} is therefore swept while the filter amplitude is kept constant. The resulting filtered signal as a function of the filter center lcfl_{c}^{f} therefore resembles the original size-distributions, where the log-normal peak and both Gaussian peaks are clearly identified. The effects of T2T_{2}-relaxation are shown in Fig. 3c for typical values of white-matter tissue. To further demonstrate the filter selectivity, Figure 3d shows that if one chooses lcfl_{c}^{f} equal to the center of one of the two Gaussian distributions, the other component is filtered-out. These simulations demonstrate the feasibility of performing quantitative images of a selective microstructure-size lcfl_{c}^{f} based on the NOGSEc amplitude as a “bandpass” filter.

Refer to caption
Figure 3: Selective size-filtering with NOGSE contrast in typical microstructure size-distributions. Size-distribution probability P(lc)P(l_{c}) for two typical cases of tissue microstructure, (a) a log-normal distribution with median 2μm2\,\mathrm{\mu m} and geometric standard deviation 1.22μm1.22\,\mathrm{\mu m}, representative of white matter size-distributions; and (b) a bimodal Gaussian distribution with means lc=2μml_{c}\!=\!2\,\mathrm{\mu m} and lc=5μml_{c}\!=\!5\,\mathrm{\mu m}, and a standard deviation 0.20.2μm\mathrm{\mu m} for both peaks. (c,d) Normalized NOGSEc amplitude ΔM\Delta M as a function of the filter center lcfl_{c}^{f}. The diffusion coefficient is D0=0.7μm2/msD_{0}\!=\!0.7\,\mu\mathrm{m^{2}/ms} in both cases. (c) ΔM\Delta M including transversal T2T_{2}-relaxation effects (dashed line, T2=80msT_{2}\!=\!80\,\mathrm{ms}) is contrasted with the ideal case without relaxation effects (solid line, T2=T_{2}\!=\!\infty) for the distribution of panel (a). Here LD2=11L_{D}^{2}\!=\!11, N=4N\!=\!4 and ΔMmax=0.08\Delta M_{max}\!=\!0.08. (d) The dashed lines show ΔM\Delta M obtained considering separately each of the component of the bimodal distribution of panel (b). The low overlap between the dashed lines demonstrates the filtering property of NOGSEc. Here LD2=25L_{D}^{2}\!=\!25, N=8N\!=\!8 and ΔMmax=0.027\Delta M_{max}\!=\!0.027.

The transversal T2T_{2}-relaxation limits the largest microstructure-size that can be filtered, as the restricted diffusion regime has to be achieved. A good SNR for ΔM\Delta M is only obtained for Lc<LD<L2L_{c}<L_{D}<L_{2}. For short T2T_{2}, the strategy is therefore using the lowest possible value of LDL_{D} and reducing the number of refocusing periods NN so as to remain in the restricted regime. We consider this scenario in proof-of-principle experiments and perform a size-filter sweep varying only the gradient amplitude while keeping TET2T_{E}\lesssim T_{2} constant. We implement the microstructure-size filter method on an ex-vivo mouse brain focusing on the Corpus Callosum (CC) region (see experimental details in Methods). The CC contains aligned axons and is a paradigmatic model for log-normal size-distributions (Assaf et al., 2008; Pajevic and Basser, 2013; White et al., 2013; Liewald et al., 2014; Shemesh et al., 2015). Figure 4a shows images of ΔM\Delta M for two gradient strengths. The largest gradient acts as a “bandpass” Gaussian-filter of the lower microscopic sizes of the distribution, compared to the weaker gradient that acts as a “high-pass” filter of the larger sizes (Fig. 4b). Therefore Fig. 4a clearly highlights zones of the CC with complementary colors depending of the microstructure-size. This is demonstrated quantitatively in Fig. 4b-c for three regions-of-interest (ROI). The average ΔM\Delta M as a function of the gradient for the ROIs is shown in Fig. 4c together with fitted curves derived from our theoretical model following Eq. (7) (see Methods). The inferred microstructure-size distributions and the ΔM\Delta M filter shapes are shown in Fig. 4b. The excellent agreement between the model and the experimental data fully demonstrates the reliability of the quantitative images shown in panel a based on the NOGSE microstructure-size filter and the assumed log-normal model.

Refer to caption
Figure 4: Non-invasive NOGSE-imaging of selective microstructure-sizes in ex-vivo mouse brain. (a) Two images based on NOGSEc of the Corpus Callosum region of a mouse brain for two gradient strengths. The ΔM\Delta M contrast highlights different zones with complementary colors when comparing the images for the two gradients. The color scale covers the full range of contrast signal in each image. Three regions-of-interest (ROI) are indicated in black contours. The images were acquired for N=2N\!=\!2 and TE=21.5msT_{E}\!=\!21.5\,\mathrm{ms}. Pixel-size 78×78μm278\times 78\mu\text{m}^{2}. (see more details in Methods). (b) Size-distributions (solid-lines) that best fit the experimental data for the selected ROIs. The dashed lines show ΔM(lc)\Delta M(l_{c}) predicted by our model for the two gradients used in (a). The predicted overlap between the reconstructed distribution and the ΔM(lc)\Delta M(l_{c}) filter is consistent with the signal contrast shown in (a and c): ROI 1 has lower sizes than ROI 2 and 3, and therefore ROI 1 has higher ΔM\Delta M for G=800mT/mG\!=\!800\,\mathrm{mT/m}; conversely ROI 2 and 3 have higher contrast amplitude for G=125mT/mG\!=\!125\,\mathrm{mT/m}.(c) Average NOGSEc signal (symbols) as a function of the gradient strength GG for the three ROIs. The vertical dashed lines mark the gradient strengths used in (a). The solid lines are fits to the experimental data of our theoretical model following Eq. (7) for a log-normal distribution. The fitted parameters are the median 1.08±0.061.08\pm 0.06, 2.82±0.012.82\pm 0.01 and 1.87±0.04μm1.87\pm 0.04\,\mu m and the geometric standard deviation 2.58±0.032.58\pm 0.03, 2.39±0.022.39\pm 0.02 and 2.91±0.04μm2.91\pm 0.04\,\mu m for ROI 1, 2 and 3, respectively. We considered a uniform D0=0.7μm2/msD_{0}\!=\!0.7\,\mu\mathrm{m^{2}/ms} as a representative diffusion coefficient to fit the size-distribution.

Conclusions. The presented results introduce a method for performing non-invasive quantitative images of selective microstructure-sizes based on probing nuclear-spin dephasing induced by molecular diffusion with magnetic resonance. Conversely to standard diffusion-weighted imaging approaches that are based on observing the decay-rate of the spin signal, we exploit dynamical control with oscillating gradients to selectively probe a decay-shift on spin-echo decays. This decay-shift contains quantitative information of microstructure-sizes that restrict the molecular diffusion. We generate a contrast amplitude that behaves as a microstructure-size filter to selectively probe a restriction length determined by the control parameters. We show the usefulness and performance of the method with proof-of-principle simulations and experiments on typical size-distributions of white-matter tracts in a mouse brain. A quantitative image of specific diffusion restriction lengths is performed extracted from only two images, allowing to significantly reduce the tens of images that typically demand the inference of microstructure-sizes from data fittings (Assaf et al., 2008; Alexander et al., 2010; Ong and Wehrli, 2010; Xu et al., 2014; Shemesh et al., 2015; Novikov et al., 2019). Even though intrinsic T2T_{2}-relaxation may represent a limitation, we show excellent performance probing quantitative information of microstructure details between 0.110μm\sim 0.1-10\,\mu\mathrm{m} on biological tissue as in the mouse white matter, being able to filter sizes much lower than the present image resolution. This work lays the foundations of a novel conceptual tool with low overhead for designing quantitative methods for non-invasive imaging of tissue microstructure. This diagnostic tool opens up a new avenue to explore for in-vivo imaging. In addition, these results can also be applied for characterizing material microstructures, such as rocks which are of particular interest for oil extraction, and for nanoscale-imaging of biological tissues with novel quantum sensors based on noise spectroscopy (Staudacher et al., 2015; Barry et al., 2016; Wang et al., 2019).

Acknowledgements.
Acknowledgments. We acknowledge Soledad Esposito and Micaela Kortsarz for preparing the ex-vivo mouse brain, and Federico Turco for scripting assistance to process the experimental data. We thank Lucio Frydman and Jorge Jovicich for fruitful discussions. This work was supported by CNEA, ANPCyT-FONCyT PICT-2017-3447, PICT-2017-3699, PICT-2018-04333, PIP-CONICET (11220170100486CO), UNCUYO SIIP Tipo I 2019-C028, Instituto Balseiro. A.Z. and G.A.A. are members of the Research Career of CONICET. M.C. and P.J. acknowledge support from the Instituto Balseiro’s fellowships.

References

Methods

Magnetization decay of an spin ensemble under dynamical control. The magnetization signal observed from an ensemble of non-interacting and equivalent spins, under the effect of dynamical control, is M(t)=eiϕ(t)M(0)M(t)=\left\langle e^{-i\phi(t)}\right\rangle M(0). Here, the brackets denote the ensemble average over the random phases ϕ(t)\phi(t) acquired by the spins during the evolution time tt. For the considered dynamical control with modulated gradient spin-echo sequences, the average phase becomes null, ϕ(t)=0\left\langle\phi(t)\right\rangle=0. Then, as ϕ(t)\phi(t) typically follows a Gaussian distribution (Stepisnik, 1999), the signal will depend on the random phase variance M(t)=e12ϕ2(t)M(0)M(t)=e^{-\frac{1}{2}\left\langle\phi^{2}(t)\right\rangle}M(0).

The variance expressed in terms of the control applied to the system by G(t)G(t) and the molecular displacement autocorrelation function Δr(0)Δr(t)=D0τce|t|/τc\left\langle\Delta r(0)\Delta r(t)\right\rangle=D_{0}\tau_{c}e^{-|t|/\tau_{c}} (Stepisnik, 1993; Álvarez et al., 2013; Shemesh et al., 2013) is given in Eq. (2) of the main text.

For a piecewise constant modulation G(t)G(t), that switches NN times its sign at times tit_{i} with i=0..N1i=0..N-1 during the evolution time TET_{E}. The quadratic phase of the magnetization decay is

ϕ2(TE)\displaystyle\left\langle\phi^{2}(T_{E})\right\rangle =γ2G2D02τci[=0]N1j[=0]N1titi+1tjtj+1e|tt|/τc(1)i(1)jdtdt\displaystyle={\displaystyle\gamma^{2}G^{2}D_{0}^{2}\tau_{c}\stackrel{{\scriptstyle[}}{{i}}=0]{N-1}{\sum}\stackrel{{\scriptstyle[}}{{j}}=0]{N-1}{\sum}\int_{t_{i}}^{t_{i+1}}\int_{t_{j}}^{t_{j+1}}}e^{-|t-t^{\prime}|/\tau_{c}}(-1)^{i}(-1)^{j}{\rm d}t^{\prime}\,{\rm d}t
=γ2G2D02τci[=0]N1j[=0]N1(1)i(1)j[({2τctjτc2etitjτctjtiτc2etitjτcti<tj)\displaystyle{\displaystyle=\gamma^{2}G^{2}D_{0}^{2}\tau_{c}\stackrel{{\scriptstyle[}}{{i}}=0]{N-1}{\sum}\stackrel{{\scriptstyle[}}{{j}}=0]{N-1}{\sum}}(-1)^{i}(-1)^{j}\left[\left(\begin{cases}2\tau_{c}t_{j}-\tau_{c}^{2}e^{-\frac{t_{i}-t_{j}}{\tau_{c}}}&t_{j}\leq t_{i}\\ -\tau_{c}^{2}e^{-\frac{t_{i}-t_{j}}{\tau_{c}}}&t_{i}<t_{j}\end{cases}\right)\right.
+({2τctj+1τc2eti+1tj+1τctj+1ti+1τc2eti+1tj+1τc+2τcti+1ti+1<tj+1)({2τctjτc2eti+1tjτctjti+1τc2eti+1tjτc+2τcti+1ti+1<tj)\displaystyle+\left(\begin{cases}2\tau_{c}t_{j+1}-\tau_{c}^{2}e^{-\frac{t_{i+1}-t_{j+1}}{\tau_{c}}}&t_{j+1}\leq t_{i+1}\\ -\tau_{c}^{2}e^{\frac{t_{i+1}-t_{j+1}}{\tau_{c}}}+2\tau_{c}t_{i+1}&t_{i+1}<t_{j+1}\end{cases}\right)-\left(\begin{cases}2\tau_{c}t_{j}-\tau_{c}^{2}e^{-\frac{t_{i+1}-t_{j}}{\tau_{c}}}&t_{j}\leq t_{i+1}\\ -\tau_{c}^{2}e^{-\frac{t_{i+1}-t_{j}}{\tau_{c}}}+2\tau_{c}t_{i+1}&t_{i+1}<t_{j}\end{cases}\right) (8)
({2τctj+1τc2etitj+1τctj+1tiτc2etitj+1τc+2τctiti<tj+1)].\displaystyle\left.-\left(\begin{cases}2\tau_{c}t_{j+1}-\tau_{c}^{2}e^{-\frac{t_{i}-t_{j+1}}{\tau_{c}}}&t_{j+1}\leq t_{i}\\ -\tau_{c}^{2}e^{\frac{t_{i}-t_{j+1}}{\tau_{c}}}+2\tau_{c}t_{i}&t_{i}<t_{j+1}\end{cases}\right)\right].

Magnetization decay within the restricted diffusion regime. In the restricted diffusion regime all terms e|titj|τc0\propto e^{-\frac{\left|t_{i}-t_{j}\right|}{\tau_{c}}}\rightarrow 0 as |titj|τc\left|t_{i}-t_{j}\right|\gg\tau_{c} for all iji\neq j, therefore the non-null terms in Eq. (8) are those with i=ji=j. The phase variance is then

12ϕ2(TE)\displaystyle-\frac{1}{2}\left\langle\phi^{2}(T_{E})\right\rangle =12γ2G2D02τci[=0]N1j[=0]N1(1)i(1)j[(2τctjτc2etitjτc)δij+(2τctj+1τc2eti+1tj+1τc)δi+1j+1\displaystyle{\displaystyle=-\frac{1}{2}{\displaystyle\gamma^{2}G^{2}D_{0}^{2}\tau_{c}\stackrel{{\scriptstyle[}}{{i}}=0]{N-1}{\sum}\stackrel{{\scriptstyle[}}{{j}}=0]{N-1}{\sum}}(-1)^{i}(-1)^{j}}\left[\left(2\tau_{c}t_{j}-\tau_{c}^{2}e^{-\frac{t_{i}-t_{j}}{\tau_{c}}}\right)\delta_{ij}+\left(2\tau_{c}t_{j+1}-\tau_{c}^{2}e^{-\frac{t_{i+1}-t_{j+1}}{\tau_{c}}}\right)\delta_{i+1j+1}\right. (9)
(2τctjτc2eti+1tjτc)δi+1j(2τctj+1τc2etitj+1τc)δij+1]\displaystyle\>\left.-\left(2\tau_{c}t_{j}-\tau_{c}^{2}e^{-\frac{t_{i+1}-t_{j}}{\tau_{c}}}\right)\delta_{i+1j}-\left(2\tau_{c}t_{j+1}-\tau_{c}^{2}e^{-\frac{t_{i}-t_{j+1}}{\tau_{c}}}\right)\delta_{ij+1}\right]
=12γ2G2D02τc[2τci[=1]N(titi1)τc2(4N+2)]\displaystyle{\displaystyle=-\frac{1}{2}{\displaystyle\gamma^{2}G^{2}D_{0}^{2}\tau_{c}}}\left[2\tau_{c}\stackrel{{\scriptstyle[}}{{i}}=1]{N}{\sum}\left(t_{i}-t_{i-1}\right)-\tau_{c}^{2}(4N+2)\right] (10)
=γ2G2D02τc2[TE(2N+1)τc].\displaystyle{\displaystyle=-\gamma^{2}G^{2}D_{0}^{2}\tau_{c}^{2}}\left[T_{E}-(2N+1)\tau_{c}\right]. (11)

Then, the decay-rate is dϕ2(TE)dTE=γ2G2D02τc2\frac{d\left\langle\phi^{2}(T_{E})\right\rangle}{dT_{E}}=\gamma^{2}G^{2}D_{0}^{2}\tau_{c}^{2}, and the decay-shift is the time-independent term γ2G2D02τc3(2N+1){\displaystyle\gamma^{2}G^{2}D_{0}^{2}\tau_{c}^{3}}(2N+1), which can also be derived from

γ2G2D0τc3(2N+1)=γ2G2D0τc(2N+1)0𝑑ttΔr(0)Δr(t).\gamma^{2}G^{2}D_{0}\tau_{c}^{3}(2N+1)\!=\!\gamma^{2}G^{2}D_{0}\tau_{c}(2N+1)\!\!\int_{0}^{\infty}\!\!\!\!dt\,t\left\langle\Delta r(0)\Delta r(t)\right\rangle. (12)

NOGSE contrast amplitude. One can obtain an analytical expression for the magnetization decay in Eq. (8) for the Hahn, CPMG and NOGSE spin-echo sequences described in Fig. (1) of the main text. In the restricted diffusion regime TE,tH,tCτcT_{E},t_{\mathrm{H}},t_{\mathrm{C}}\gg\tau_{c}, they result

MHahn(tH)=exp{γ2G2D0τc3[tHτc3]},MCPMG(NtC,N)=exp{γ2G2D0τc3[NtCτc(2N+1)]},MNOGSE(TE,N,tC)=exp{γ2G2D0τc3[TEτc(2N+1)]},\begin{split}&M_{\mathrm{Hahn}}(t_{\mathrm{H}})=\mathrm{exp}\{-\gamma^{2}G^{2}D_{0}\tau_{c}^{3}[\frac{t_{\mathrm{H}}}{\tau_{c}}-3]\},\\ &M_{\mathrm{CPMG}}(Nt_{\mathrm{C}},N)=\mathrm{exp}\{-\gamma^{2}G^{2}D_{0}\tau_{c}^{3}[\frac{Nt_{\mathrm{C}}}{\tau_{c}}-(2N+1)]\},\\ &M_{NOGSE}(T_{E},N,t_{C})=\mathrm{exp}\{-\gamma^{2}G^{2}D_{0}\tau_{c}^{3}[\frac{T_{E}}{\tau_{c}}-(2N+1)]\},\end{split} (13)

with TE=(N1)tC+tHT_{E}=(N-1)t_{\mathrm{C}}+t_{H}.

We define the NOGSE contrast (NOGSEc) amplitude ΔM\Delta M to the difference between MNOGSE(TE,N,tC=tH)=MCPMG(TEN,N)M_{NOGSE}(T_{E},N,t_{C}=t_{H})=M_{\mathrm{CPMG}}(\frac{T_{E}}{N},N) and MNOGSE(TE,N,tC0)MHahn(TE)M_{NOGSE}(T_{E},N,t_{C}\rightarrow 0)\simeq M_{\mathrm{Hahn}}(T_{E})), i.e.

ΔM(TE,N)=MCPMG(TE,N)MHahn(TE).\Delta M(T_{E},N)=\mathrm{M_{\mathrm{CPMG}}(T_{E},N)-M_{\mathrm{Hahn}}(T_{E})}. (14)

Then, we arrive to Eq. (3) of the main text by introducing Eq. (13) into Eq. (14)

ΔMeγ2G2D0τc3(TEτc3)(eγ2G2D0τc32(N1)1).\Delta M\approx e^{-\gamma^{2}G^{2}D_{0}\tau_{c}^{3}(\frac{T_{E}}{\tau_{c}}-3)}\!\left(e^{\gamma^{2}G^{2}D_{0}\tau_{c}^{3}2(N-1)}-1\right). (15)

NOGSEc results

ΔM(LD,Lc,N)=eLc4(LD23Lc2)(e2(N1)Lc61),\Delta M(L_{D},L_{c},N)=e^{-L_{c}^{4}(L_{D}^{2}-3L_{c}^{2})}(e^{2(N-1)L_{c}^{6}}-1), (16)

within the restricted diffusion regime, using the dimensionless variables LD,LcL_{D},L_{c} defined in the main text.

The general expression for ΔM\Delta M that includes all diffusion time scales can be obtained from Eq. (8) by replacing the time intervals as defined in Fig. 1c of the main text.

Gaussian microstructure-size filter derivation. The NOGSEc amplitude in the restricted diffusion regime, Eq. (16), can be approximated by

ΔMeLc4(LD23Lc2)2(N1)Lc6,\Delta M\approx e^{-L_{c}^{4}(L_{D}^{2}-3L_{c}^{2})}2(N-1)L_{c}^{6}, (17)

for Lc1L_{c}\ll 1. The maximum of ΔM\Delta M occurs at dΔMdLC=0\frac{d\Delta M}{dL_{C}}=0. In the asymptotic limit of LD1L_{D}\gg 1, it is achieved for

Lc=Lcf(3/2)14LD12+𝒪(LD72),L_{c}=L_{c}^{f}\approx\left(3/2\right)^{\frac{1}{4}}L_{D}^{-\frac{1}{2}}+\mathcal{O}(L_{D}^{-\frac{7}{2}}), (18)

where LcfL_{c}^{f} is the center of the filter as described in Eq. (5) of the main text.

The NOGSEc amplitude can be approximated by

ΔM\displaystyle\Delta M \displaystyle\approx e3/2332(N1)LD3\displaystyle{\rm e}^{-3/2}3\sqrt{\frac{3}{2}}\left(N-1\right)L_{D}^{-3}
36e3/2(N1)LD2(LcLcf)2+𝒪((LcLcf)4),\displaystyle-36{\rm e}^{-3/2}\left(N-1\right)L_{D}^{-2}(L_{c}-L_{c}^{f})^{2}\!+\!\mathcal{O}((L_{c}-L_{c}^{f})^{4}),

with a Taylor expansion in LcL_{c} at LcLcfL_{c}\approx L_{c}^{f} of the expression given in Eq. (17). We use this expansion to define the first moments of the Gaussian filter function of Eq. (4) in the main text, obtaining

ΔM2(N1)e3/2(Lcf)6exp[12(LcLcfLcf)2].\Delta M\approx 2(N-1)e^{-3/2}\left(L_{c}^{f}\right)^{6}\exp\left[-12\left(\frac{L_{c}-L_{c}^{f}}{L_{c}^{f}}\right)^{2}\right]. (20)

This expression is then verified to approximate very well the exact expression derived from Eq. (8) within the regime of LD1L_{D}\gg 1 and Lc1L_{c}\ll 1.

Ex-vivo mouse brain preparation. The experiments were approved by the Institutional Animal Care and Use Committee of the Comisión Nacional de Energía Atómica under protocol number 08_2018. One mouse was sacrificed by isoflurane overdose and its brain was fixed in formaline. The brain was washed twice with PBS prior to the insertion into a 15 ml falcon tube filled with PBS. The brain was left in the magnet for at least three hours prior to the reported experiments to reach thermal equilibration.

MRI experiments. The experiments were performed on a 9.4T Bruker Avance III HD WB NMR spectrometer with a 1H resonance frequency of ωz=400.15\omega_{z}=400.15 MHz. We use a Micro 2.5 probe capable of producing gradients up to 1500 mT/m in three spatial directions. The experiments temperature was stabilized at 21°C. We programmed and implemented with Paravision 6 the NOGSE MRI sequence shown in Fig. 5. The sequence parameters were: Repetition time 20002000 ms, Techo time=55msT_{\text{echo time}}=55\text{ms}, FOV = 15x15 mm2 with a matrix size of 192x192, leading to an in-plane resolution of 78×78μm278\times 78\mu\text{m}^{2}, and slice thickness of 1 mm with 128 signal averages. The two images were acquired with echo planar imaging (EPI) encoding with 4 segments (image acquisition time 17\approx 17 min) and then subtracted to generate ΔM\Delta M. The NOGSE modulation time was TE=21.5msT_{E}=21.5\text{ms} with N=2N=2. The NOGSE gradients were applied perpendicular to the main axis of the axons in the corpus callosum. NOGSEc ΔM\Delta M is determined from an image generated with tH=tC=10.75mst_{H}=t_{C}=10.75\text{ms} for the CPMG modulation and with tH=0.5mst_{H}=0.5\text{ms} and tC=21mst_{C}=21\text{ms} for the Hahn modulation. The set of parameter values were chosen for achieving good SNR for performing the proof-of-principle experiments. Further studies should be considered to explore the optimal values for acquiring the images in the shortest possible time.

Refer to caption
Figure 5: Scheme for the experimental implementation of the NOGSE sequence. An initial selective RF-excitation π2\frac{\pi}{2}-pulse is applied to select a tissue slice. It is followed by a NOGSE gradient modulation of duration TET_{E} following the scheme described in Fig. 1d of the main text. During the evolution time TET_{E}, the gradient strength and sign is modulated with trapezoidal shapes. Then a selective RF π\pi-pulse is applied to refocus magnetic field inhomogeneities. At the end a spatial EPI-enconding is applied for acquiring an image. Three gradients are applied in the three spatial directions for slide selection GslG_{sl}, for read orientation GROG_{RO} and phase encoding GPEG_{PE}. The NOGSE gradients can be applied in arbitrary orientations.

Experimental data analysis. The mean signal from the pixels in the ROIs of Fig. 4a of the main text was analyzed, and plotted as a function of GG in Fig. 4c. Fittings to the theoretical model were done assuming a uniform D0=0.7μm2/msD_{0}=0.7\mu m^{2}/\text{ms} and a log-normal distribution P(lc)=12πσlce(ln(lc)μ)22σ2P(l_{c})=\frac{1}{\sqrt{2\pi}\sigma l_{c}}e^{-\frac{(ln(l_{c})-\mu)^{2}}{2\sigma^{2}}} with median eμe^{\mu} and geometric standard deviation eσe^{\sigma}. This implies that no extra assumptions were considered for the tissue model (e.g., intra/extra-cellular compartments). Therefore a single log-normal distribution was thus fitted to the experimental data, regardless of the potential heterogeneity. This means that all underlying compartments (e.g., extracellular, intracellular, etc.) reflected in the diffusion weighted are assumed to be described by a single log-normal distribution. We considered a distribution of restriction lengths lcl_{c} without assuming particular geometries. Remarkably the excellent agreement of the fitted curves to the experimental data in Fig. 4c is consistent with these simple assumptions.