Demonstration of the nearly continuous operation of an 171Yb optical lattice clock for half a year
Abstract
Optical lattice clocks surpass primary Cs microwave clocks in frequency stability and accuracy, and are promising candidates for a redefinition of the second in the International System of Units (SI). However, the robustness of optical lattice clocks has not yet reached a level comparable to that of Cs fountain clocks which contribute to International Atomic Time (TAI) by the nearly continuous operation. In this paper, we report the long-term operation of an 171Yb optical lattice clock with a coverage of for half a year including uptimes of for the first 24 days and for the last 35 days. This enables a nearly dead-time-free frequency comparison of the optical lattice clock with TAI over months, which provides a link to the SI second with an uncertainty of low . By using this link, the absolute frequency of the 1SP0 clock transition of 171Yb is measured as Hz with a fractional uncertainty of . This value is in agreement with the recommended frequency of 171Yb as a secondary representation of the second.
1 Introduction
Several optical lattice clocks have achieved fractional frequency stabilities and uncertainties at the level [1, 2, 3, 4, 5], which is better than primary Cs fountain microwave clocks. These achievements have stimulated discussion regarding a redefinition of the second in the International System of Units (SI) [6, 7, 8, 9].
With the aim of redefining the SI second, the absolute frequencies of optical lattice clocks have been measured by many groups and used to determine the recommended frequencies of secondary representations of the second [8]. The absolute frequency can be directly measured referenced to a Cs fountain clock if it is locally available in the laboratory [14, 11, 10, 12, 13, 15]. When a local Cs clock is not available, the absolute frequency can be measured using a satellite link to International Atomic Time (TAI). TAI is a global timescale computed by the Bureau International des Poids et Mesures (BIPM). BIPM provides the frequency difference between TAI and the SI second averaged over a month. Therefore, the continuous comparison of an optical clock with TAI for a month is desirable in terms of linking to the SI second without additional uncertainty resulting from the dead time. A long-term comparison is also needed to reduce the uncertainty arising from the satellite link [16]. Since optical lattice clocks have commonly been operated intermittently, previous works have mostly employed stable local flywheels (e.g, an ensemble of hydrogen masers) that reach the level to bridge the gaps in the operation of the lattice clocks [17, 18, 19, 20]. A few groups have reported the nearly continuous operation of Sr optical lattice clocks with uptimes of for 10 days, for three weeks [14], and for 25 days [21].
The continuous operation of an optical lattice clock is essential for the calibration of the frequency of TAI as a secondary representation of the second [8] as well as the absolute frequency measurement. A robust lattice clock is also important for new applications including tests of fundamental physics [12, 19, 22, 23, 24, 25, 26], relativistic geodesy [27, 28], and the generation of a stable local timescale [13, 29, 30, 31].
The main challenge as regards realizing the continuous operation of an optical lattice clock is to stabilize the frequencies of several light sources including an optical frequency comb to allow the comparison of the optical clock and a microwave standard. So far, we have developed multi-branch erbium-doped-fiber-based frequency combs that can run continuously for long periods [32]. By stabilizing the frequencies of many light sources to the comb, we have constructed a simple and reliable laser system for clock experiments [33]. We have also developed a relocking scheme [34] for optical phase locking. These techniques should greatly assist the realization of the continuous operation.
In this paper, we report the operation of an 171Yb optical lattice clock with a coverage of for half a year (185 days) including uptimes of for the first 24 days, for the second 27 days, for the third 30 days, for the fourth 35 days, for the fifth 25 days, and for the sixth 35 days. Using a single hydrogen maser with its stability limited by a flicker floor of , we reduce the uncertainty in the link between the Yb clock and the SI second to the low level, and measure the absolute frequency of the Yb clock transition.
2 Experimental setup

Figure 1 is a schematic diagram of the experimental setup. The setup for the Yb optical lattice clock is similar to that reported in our previous paper [35]. The clock utilizes the 1SP0 clock transition at 578 nm of 171Yb atoms trapped in an optical lattice operated at the magic wavelength [36] of 759 nm. The 578 nm light was generated by the second harmonic generation (SHG) of an external cavity diode laser (ECDL) emitting at 1156 nm [37]. The lattice light at 759 nm was provided by a titanium sapphire (Ti:S) laser (M squared, SolsTiS-4000-SRX-R). Before loading the atoms to the optical lattice, two-stage laser cooling with the 1SP1 transition at 399 nm and the 1SP1 transition at 556 nm was carried out to decrease the temperature of the atoms to several tens of microkelvins. The 399 nm light was generated by an injection-locked 399 nm diode laser with an automatic relocking mechanism [38]. A seed light for the injection locking was provided by the SHG of a 798 nm ECDL [39]. The 556 nm light for the second stage cooling was prepared by the SHG of a 1112 nm light from a fiber laser (NKT Photonics, Koheras) with a semiconductor optical amplifier. All the SHG processes were performed with single-pass periodically poled LiNbO3 waveguides.
The 1112 nm fiber laser and the 1156 nm ECDL were phase locked to a narrow linewidth fiber comb (Fiber comb 1 in Fig. 1) [41, 40]. This comb was phase locked to a master Nd:YAG laser at 1064 nm, which was stabilized to an ultra-low expansion (ULE) cavity by using the Pound-Drever-Hall method. To compensate for the narrow capture range of optical phase locking, a recently-developed relocking scheme [34] was incorporated in the phase locking of the comb to the master laser and the phase locking of the 1112 nm and 1156 nm lasers to the comb. The 798 nm ECDL and the Ti:S laser were stabilized to another fiber comb (Fiber comb 2 in Fig. 1) by frequency locking with an electrical delay line [42], which ensured a large capture range. Fiber comb 2 was referenced to the Coordinated Universal Time of the National Metrology Institute of Japan (UTC(NMIJ)), which is generated by a hydrogen maser and an auxiliary output generator to add an arbitrary frequency offset.
The Yb optical lattice clock was operated with a cycle time of about 1.7 s. Yb atoms effused from an oven were decelerated with a Zeeman slower and cooled in first-stage cooling for about 1.2 s and then further cooled in second-stage cooling for 275 ms. In the present work, a window heated at 204∘C was installed inside a vacuum chamber in the pass of the Zeeman slower beam to prevent the atomic beam from coating the viewport. The atoms were then loaded into a vertically oriented one-dimensional optical lattice and spin-polarized with the 1SP1 transition at 556 nm for 20 ms. A Rabi pulse resonant on the clock transition was then applied for 40 ms. The atomic population and the excitation probability were deduced using a laser-induced fluorescence signal by the 1SP1 transition at 399 nm, and repumping on the 3PD1 transition at 1389 nm induced by a free-running distributed feedback laser. The excitation probability was used to calculate frequency corrections applied for an acousto-optic modulator (AOM), which steers the frequency of the clock laser to the atomic transition. The atomic transition was split into two Zeeman components by applying a bias magnetic field of 65 T. The clock laser was stabilized to each component.
Fiber comb 2 was also used to measure the frequency of the Yb clock against UTC(NMIJ). To transfer the stability of UTC(NMIJ) to the comb without degradation, we mixed the 7th harmonics of the repetition rate frequency MHz with a 900 MHz signal which was generated by the 90-fold frequency multiplication of a 10 MHz signal from UTC(NMIJ). The resulting output frequency from the mixer MHz was phase locked to a synthesizer referenced to UTC(NMIJ). After stabilizing the comb to UTC(NMIJ), the beat frequency between the clock laser and the comb was measured by using a zero dead-time frequency counter.
3 Demonstration of the nearly continuous operation
The frequency of the Yb optical lattice clock was measured against UTC(NMIJ) during a half-year period from the Modified Julian Date (MJD) 58754 (28 September 2019) to MJD 58939 (31 March 2020). The Yb clock was operated with uptimes of 93.9 in the first 24-day campaign from MJD 58754 to MJD 58778, 86.4 in the second 27-day campaign from MJD 58787 to MJD 58814, 80.4 in the third 30-day campaign from MJD 58814 to MJD 58844, 72.7 in the fourth 35-day campaign from MJD 58844 to MJD 58879, 82.6 in the fifth 25-day campaign from MJD 58879 to MJD 58904, and 92.6 in the sixth 35-day campaign from MJD 58904 to MJD 58939 (see also Table 1). Figures 2 (a) - (f) respectively show the fractional frequency differences between the Yb clock and UTC(NMIJ), denoted by , in the first - sixth campaigns. The uptime was calculated by using , where is the number of valid data points, the clock cycle time ( s), and the total period of the measurement. We discarded data that included the following events: (i) large excursions in the phase-locked frequencies that can occur during the relocking procedures; (ii) cycle slips in the frequency counting; and (iii) low excitation probabilities of the clock transition that lasted for a sufficiently long period.
The fractional frequency difference is given by
(1) | |||||
where denotes the actual (nominal) frequency of X. The approximation is vaild when . Here we chose MHz and Hz, which is the CIPM (Comité International des Poids et Mesures) recommended frequency of 171Yb [8].
Campaign | MJD | Period (days) | Uptime |
---|---|---|---|
1 | 58754 - 58778 | 24 | |
2 | 58787 - 58814 | 27 | |
3 | 58814 - 58844 | 30 | |
4 | 58844 - 58879 | 35 | |
5 | 58879 - 58904 | 25 | |
6 | 58904 - 58939 | 35 | |
Total | 58754 - 58939 | 185 |

The clock was mostly unattended during the full campaign period. Several experimental parameters were monitored remotely. A data acquisition computer automatically sent an email alert to operators when one of the experimental parameters was outside the normal range. In the first campaign (see Fig. 2 (a)), to demonstrate almost dead-time-free operation ( uptime per day), we always carried out manual tuning as soon as we received the email alert. Significant reductions in the uptime were caused by rare events such as the passage of a typhoon at MJD 58770 and the breaking of a diode laser used in fiber comb 1 at MJD 58778. In the second to sixth campaigns (see Figs. 2 (b)-(f)), manual recovery was not always performed at night to demonstrate that a large uptime can also be achieved with minimum human effort. Large gaps from MJD 58831 to MJD 58834 in the third campaign were caused by the shutdown of the air conditioners for facility maintenance. In the fourth campaign, a diode laser used in fiber comb 2 was broken, which caused a large downtime from MJD 58873 to MJD 58875. At MJD 58877, we stopped the clock operation and carried out a systematic evaluation of the microwave synthesis (See Section 5).
Figure 3 shows the distribution of interrupting events that occurred during the full campaign period. The number of interrupting events were counted according to the email alert history. We could basically link the email history to the types of interruptions, but could not link some events due to a lack of records. These events are indicated by “No record” in Fig. 3. The clock operation was mostly interrupted by acoustic noise from a drainage pump in the facility ( of the total number of interruptions), the instability of the injection locking (), mode hopping of the Ti:S laser () and the 1156 nm ECDL (), and a small number of trapped atoms (). Earthquakes () stopped the Yb clock mostly by dropping the lock of the 798 nm diode laser without implementing the relocking scheme. The other relockable light sources were resistant to small magnitude earthquakes. The other factors that interrupted the operation included rapid frequency drifts of the ULE cavity after the air conditioners were shutdown (MJD 58831 to MJD 58834) and temporary freezing of the data acquisition computer.


Figure 4 shows the number of interrupting events in each campaign. Most of the interruptions occurred with a small number () in each campaign except for the Ti:S laser (blue square), a small number of trapped atoms (green square), and a drainage pump (red square). The lock of the Ti:S laser failed frequently especially in the second campaign. This problem was partially solved in the following campaigns by cleaning the mirrors used to introduce the pump laser beam and improving the signal-to-noise ratio of the error signal for the frequency locking to the comb. From the second campaign, the number of atoms trapped in the optical lattice gradually decreased (see Section 4), which increased the occurrence of lock failures. From the end of the third campaign, the Yb clock was frequently interrupted by a drainage pump, which had been installed in the facility during the third campaign. The pump generated acoustic noise for about 30 minutes once a day when it drained water. During this period, the excitation probability of the atomic transition became almost zero due to the broadening of the clock laser linewidth.
4 Frequency stability and uncertainty of the optical lattice clock

Effect | Shift | Uncertainty |
---|---|---|
() | () | |
Lattice light | 3.4 | 33.1 |
BBR | 20.8 | |
Density | 6.4 | |
Second-order Zeeman | 0.3 | |
Probe light | 0.4 | 0.2 |
Servo error | 1.1 | |
AOM switching | 1 | |
Total | 39.6 |

We evaluated the frequency stability of the Yb optical lattice clock by comparing it with our Sr optical lattice clock [43]. Figure 5 shows the Allan deviation of the frequency ratio Yb/Sr, which was improved with a slope of and reached low in several thousand seconds. Figure 5 also shows the frequency stability of UTC(NMIJ) relative to the Yb clock, which was calculated from the data of the first campaign (see Fig. 2 (a)). During this campaign period, the frequency offset of the auxiliary output generator for UTC(NMIJ) was constant.
Table 2 lists the systematic shifts and uncertainties of the Yb optical lattice clock obtained in the first campaign. The uncertainty evaluation is mostly based on our previous evaluation [35]. The density shift and the servo error differed slightly depending on the data obtained in each campaign.
The lattice induced light shift was estimated using a model [44, 45, 46] that includes the contributions from the dominant electric-dipole () polarizability, the multipolar ( and ) polarizability, and hyperpolarizability [47]. To calculate the shift using this model, the effective trap potential depth , which is less than the maximum trap depth due to the radial motion of the atoms, and the average vibrational quantum number were estimated by sideband spectroscopy of the clock transition [48]. The obtained values were and . The frequency of the lattice laser was stabilized at MHz.
To find the polarizability coefficient and the magic frequency , we have previously measured the light shift as a function of the frequency of the lattice laser [35], and fitted the obtained data with the light shift model [45]. To estimate the variation of and resulting from the systematic uncertainty of the fixed parameters in the light shift model (, , and the multipolar and hyperpolarizability shifts), we have employed a Monte Carlo method in which the fittings were repeated with different permutations of the fixed parameters. Since the distribution of obtained from many fittings was asymmetric and had a very long tail mainly due to a relatively large uncertainty in our estimation of , we have previously calculated a mean value of by removing rare events in the tail. Here, we employed a value of that was determined with center values of the fixed parameters. The uncertainty of was estimated from a coverage interval of the distribution of . The same method was applied to the estimation of . We obtained mHz/(MHz) and MHz, which agreed with our previous estimation [35] and the values reported by other groups [46, 47, 49, 20]. Finally, the light shift was estimated to be with , , , , estimated here, and the sensitively coefficients for the multipolar and hyperpolarizability shifts used in our previous evaluation [35].
The blackbody radiation (BBR) shift was estimated to be , which includes the contribution from the vacuum chamber and the atomic oven considered in our previous evaluation [35] and also that from the newly-installed heated window at 204∘C. To include this new contribution, the BBR shift induced by the heated window was calculated using a model [50] in which the atoms are located at the center of a stainless-steel sphere with an emissivity of 0.1. The radius of the sphere is equal to the distance between the atoms and the heated window ( mm). The BBR photons at 204∘C are provided from a small portion of the sphere’s surface with its area equal to that of the heated window seen by the atoms ( mm2). This model takes into account the diffuse reflection of the BBR photons on the stainless-steel wall, which increases the effective solid angle by a factor of compared with the geometric solid angle. With the sensitivity coefficient [51, 52], the calculated shift due to the heated window was . This shift value was also conservatively taken as an uncertainty.
The density shift was estimated from the atomic density in each campaign. The blue points in Fig. 6 show the atomic density data. The density decreased with a time scale of days largely due to a reduction in the transmission efficiency of the 399 nm light through the optical fiber and the output power of the 1112 nm semiconductor amplifier. Some of these components were replaced (see the caption of Fig. 6) when we could not keep the stabilization of the clock laser to the atomic transition. The status of the stabilization was confirmed by monitoring the Zeeman splittings and the excitation probabilities (see pink and green points in Fig. 6). The short-term (a few days) fluctuation of the density was mainly caused by variations in the spectral purity of the injection locking, the polarization rotation of the cooling lasers through the optical fiber, and pointing instabilities of the cooling beam at 399 nm induced by an AOM or a damaged optical fiber. In the first campaign (see blue points in Fig. 6 (a)), the mean value of the atomic density was with m-3. The uncertainty of was estimated from the standard deviation. We also added an uncertainty of to the estimation of the density according to the fluctuation of the 399 nm probe power. With our sensitivity coefficient [35] and a scaling of the trap volume with [3], the shifts were determined as , , , , , and in the first, second, third, fourth, fifth, and sixth campaigns, respectively. We observed that the mean excitation probability changed from 0.42 to 0.50 after the air conditioners were shutdown (MJD 58831 to MJD 58834) (see Fig. 6 (c)). We attributed this to a variation of the spin-polarization efficiency due to a large frequency change of the ULE cavity. This can change the shift coefficient, since the density shift depends on (i) the excitation probability mostly due to the wave interaction [53, 54, 55] and (ii) the impurity of the spin-polarization [54, 4]. Based on the measured results of Ref. [54], we estimated that in our relatively low atomic density, the change due to those effects is negligibly small compared with the uncertainty in our evaluation.
The second-order Zeeman shift was estimated from the Zeeman splitting measured in each campaign (see pink points in Fig. 6). In the first campaign, the mean value of the splitting was 268(3) Hz, where the uncertainty was estimated from the standard deviation. The shift was found to be using our sensitivity coefficient [35]. The mean splitting value was almost constant (268 - 270 Hz) during the entire campaign period.
The probe light shift was calculated to be from the shift value of Ref. [4] and the fact that the laser intensity adjusted for the pulse was inversely proportional to the time duration of the pulse.
The servo error was estimated by averaging the differences between the excitation probabilities of the high- and low-spectral shoulders for the data obtained in each measurement campaign. The servo errors were estimated to be , , , , , and in the first, second, third, fourth, fifth, and sixth campaigns, respectively. The relatively large uncertainty of in the third campaign arose from the rapid frequency drifts of the ULE cavity after the air conditioners were shutdown (MJD 58831 to MJD 58834).
5 Uncertainty of the frequency comparison by the comb

To evaluate the uncertainty of the comb-based frequency comparison between the Yb clock and UTC(NMIJ), we compared fiber comb 2 (see Fig. 1) with another comb (fiber comb 3, not shown in Fig. 1) that was independent except for the 10 MHz reference signal of UTC(NMIJ). These combs had independent setups for the microwave synthesis including the 90-fold frequency multiplication. We simultaneously measured the frequency of an acetylene-stabilized laser at 1.5 m against UTC(NMIJ) by using these two combs, and calculated the frequency difference between the combs. In the first measurement from MJD 58712 to MJD 58713, the fractional frequency difference (fiber comb 2 - fiber comb 3) was found to be , which was the average of the frequency difference data obtained for s. Figure 7 shows the Allan deviation calculated from the frequency difference data. The Allan deviation is likely limited by a flicker floor of low at averaging times s. By fitting the Allan deviation with a function which includes the white phase, the white frequency, and the flicker frequency noise components, the flicker component was estimated to be . In the second measurement at MJD 58877, the frequency difference and the flicker component were and , respectively. We took an average of the two flicker noise values () and divided it by . The obtained value was adopted as an uncertainty of the frequency comparison. Previous optical-optical comparisons [56] have shown that the uncertainty resulting from the comb itself is negligibly small. We also carried out another measurement in which the two combs shared the common frequency multiplier, and observed that the Allan deviation was improved with a slope for s. Thus, we attributed the flicker noise mainly to the frequency multiplier.
6 Absolute frequency measurement
Since UTC(NMIJ) was continuously compared with TAI via a satellite link (see Fig. 1), the frequency of the Yb clock referenced to the SI second can be deduced from the data in Fig. 2 by the relationship,
(2) | |||||
where and denote the fractional frequency differences between (i) UTC(NMIJ) and TAI and (ii) TAI and SI, respectively. These values are provided in Circular T [57], which is a monthly report issued by BIPM. Since BIPM only computes for 5-day intervals, we chose a 25-day period (MJD 58754 - MJD 58779), a 30-day period (MJD 58784 - MJD 58814), a 30-day period (MJD 58814 - MJD 58844), a 35-day period (MJD 58844 - MJD 58879), a 20-day period (MJD 58879 - MJD 58884 and MJD 58899 - MJD 58904), and a 35-day period (MJD 58904 - MJD 58939) for the first - sixth campaigns, respectively, to calculate the average value of . These periods are indicated by in Fig. 2. In the fifth campaign, a 5-day period from MJD 58894 to MJD 58899 (see the shaded region of Fig. 2 (e)) was excluded, since a large excursion of the phase of UTC(NMIJ) occurred during maintenance of the hydrogen maser at MJD 58896. Figure 8 shows the value obtained for each campaign.

Effect | First campaign | Full period |
---|---|---|
() | () | |
Statistics | ||
Yb | ||
Dead time in Yb - UTC(NMIJ) | 2.0 | 0.6 |
Dead time in TAI - SI | 1.6 | 0.4 |
UTC(NMJI) - TAI satellite link | 2.3 | 0.8 |
TAI - SI | 0.9 | 0.4 |
Statistics total | 3.6 | 1.8 |
Systematics | ||
Yb | ||
Microwave synthesis | 2.2 | 2.2 |
TAI - SI | 1.1 | 1.1 |
Gravitational red shift | 0.6 | 0.6 |
Systematics total | 4.7 | 4.7 |
Total | 5.9 | 5.0 |
The choice of the period causes an uncertainty due to the dead time in the comparison of the Yb clock and UTC(NMIJ). This dead time uncertainty was calculated by using a numerical simulation described in Ref. [58]. The simulation generated time series data of the frequency according to the noise characteristics: for the white phase modulation, for the white frequency modulation (FM), for the flicker FM, and for the random walk FM. These noise parameters were chosen to reproduce the typical stability of UTC(NMIJ) (see green curve in Fig. 5). We generated two hundred time series patterns and calculated the standard deviation between the frequency averaged over the entire 25-, 30-, or 35-day period and the mean frequency for the time during which the Yb clock was operated. This standard deviation corresponds to the dead time uncertainty. The dead time uncertainties were estimated as , , , , , and for the first, second, third, fourth, fifth, and sixth campaigns, respectively. A frequency correction arose from the dead time when the frequency steering of UTC(NMIJ) was carried out. We added corrections of , , and in the third, fifth, and sixth campaigns, respectively.
In addition, the uncertainty resulting from the dead time when we compare TAI and SI arises from the fact that BIPM calculates the value based on an evaluation with primary and secondary frequency standards for 25, 30, or 35 days, whereas our measurement period did not fully cover the evaluation period of BIPM in the first and fifth campaigns. This uncertainty was estimated to be in the first campaign, and in the full period based on the stability of the Echelle Atomique Libre (EAL) given in Circular T: for the white FM, for the flicker FM, and for the random walk FM.
The satellite link uncertainty in was calculated to be for days, for days, and for days using the recommended formula [16] with the statistical uncertainty ns reported in Circular T.
The statistical and systematic uncertainties of were calculated from the total uncertainty of and the systematic uncertainties of the primary and secondary frequency standards reported in Circular T. During the present half-year period, the frequency of TAI was calibrated by using the following frequency standards: Cs thermal beam clocks (PTB-Cs1 and PTB-Cs2 [59]), Cs fountain clocks (SYRTE-FO1, SYRTE-FO2, SYRTE-FOM [60], PTB-CSF1, PTB-CSF2 [61], SU-CsFO2 [62], and NIM5 [63]), a Rb fountain clock (SYRTE-FORb [64]), and a Sr optical lattice clock (NICT-Sr1 [29]). Since Circular T only provides the total uncertainty of , the systematic uncertainty of was estimated using the systematic uncertainties and weights of the individual frequency standards based on a method described in Ref. [65]. The estimated systematic uncertainty of changed between and during the half year period. We adopted an average value of for the full campaign period. The statistical uncertainty of was calculated with the above systematic uncertainty to reproduce the total uncertainty given in Circular T. The statistical uncertainty ranged between and , and decreased to in the full campaign period.

Table 3 summarizes the uncertainties for the absolute frequency measurement. The statistical uncertainty of the Yb clock was calculated using the stability of the Yb/Sr ratio measurement (see Fig. 5) with the duration of the operation. The gravitational red shift was estimated to be using the Yb height from the conventionally adopted geoid potential (62 636 856.0 m2/s2). The total statistical uncertainty in the full period was (dashed blue lines in Fig. 8), which was obtained by inflating the uncertainty of the weighted mean by the square root of the reduced chi-square . Finally, we obtained . The absolute frequency of the Yb clock transition was determined as Hz. The total uncertainty was mostly limited by the systematic uncertainties of the Yb clock and the microwave synthesis. Our determined value agreed with the recommended frequency and the results of previous measurements by many groups (see Fig. 9).
7 Discussions and conclusion
The uncertainty in the link between the Yb clock and the SI second was , which was calculated by using the quadratic sum of the uncertainties due to the dead times in both Yb - UTC(NMIJ) and TAI - SI, the satellite link, and the microwave synthesis. The reduction of the link uncertainty by the continuous operation also enables measurements of the frequency ratios at the level between remotely-located secondary representations of the second. For example, during the present half-year period, a Rb fountain clock (SYRTE FORb) [64] was running. The frequency difference data for 180 days were obtained from Circular T. With our data , we determined the Yb/Rb ratio as with a relative uncertainty of . This agreed with the ratio calculated from the recommended frequencies [8] and the value reported by another group [19]. As another example, a Sr optical lattice clock (NICT-Sr1 [29]) calibrated the TAI frequency for 20 days during the sixth campaign period. This enabled us to determine the Yb/Sr ratio as with a fractional uncertainty of . This was in agreement with the most accurate previous value [45] and the other values with uncertainties at the level [28, 70, 71, 20].
The expectation value of the uptime of the Yb clock was calculated under the condition where the Yb clock is fully unattended during a night interval of 8 hours, and where long downtimes resulting from rare events such as facility maintenance are not included. The number of interrupting events that required manual recovery was during the full campaign period. This was obtained by counting the number of time gaps of more than 1 minute in the frequency data (Fig. 2). Since the Yb clock was operated with a uptime for 185 days, the mean number of interruptions per hour was /hour. To estimate the probability of the occurrence of interrupting events in an interval of hours, we employed a Possison distribution,
(3) |
To calculate the expected uptime, the typical time for manually restarting the operation is needed. Figure 10 shows a histogram of the recovery time, which was obtained from time gaps in the frequency data (Fig. 2). Excluding long gaps exceeding half a day, the mean recovery time was hours. The expected dead time during a 16-hour period of human work was then calculated as hours. During an 8-hour night period, the operation is stopped after one interrupting event. In this case, the expected dead time is calculated as hours. The uptime per day was therefore expected to be .

To consider the optimum conditions for the operation, we carried out a simulation of the dead time uncertainty as a function of the clock uptime (see Figure 11). In this simulation, the dead time was homogeneously distributed for a 30-day period, except for a large time gap of days in the middle of the period. Without the large gap (indicated as “None” in Fig. 11), the dead time uncertainty falls below for uptimes of . When the clock operation is stopped for days, the uncertainty no longer reaches below even though the uptime for the overall period is . This indicates that the uncertainties calculated with the actual duration of the clock operation are mostly determined by the instabilities of UTC(NMIJ) during the large gaps. When we take into account the fact that the uncertainty of the absolute frequency measurement is limited by the systematic uncertainties of the Yb clock and the microwave synthesis (the combined uncertainty is ), it may be enough to reduce the dead time uncertainty to . In this case, the optimum condition is to operate the clock every day with a moderate uptime of . This is well below our expectation value of the uptime (, see above). It should be noted that daily repetition of the startup and shutdown of the clock increases human effort, and thus it is better in practice to let the clock run continuously.

Previous studies have demonstrated an improvement in local timescales by steering flywheel oscillators to intermittently operated lattice clocks [13, 29, 30, 31]. In these methods, the stabilities of the timescales are limited by the fluctuations of the flywheels during the dead time of the clocks, and thus can be further improved by continuous operation. An analysis for our local timescale will be presented in a future publication.
The performance at the present level can make a significant contribution to the search for new physics. For example, local position invariance has been tested by looking for an annual variation of the gravitational red shift of atomic clocks [22, 12, 19]. As demonstrated in previous experiments that utilized continuously running fountain clocks and hydrogen masers [73], the continuous operation contributes to reduce the statistical uncertainty in a relatively short term, resulting in a stringent limit on violation of local position invariance. In our case, this experiment can be carried out by comparing the Yb clock with a local Cs fountain clock [74] or a hydrogen maser. Another example is that transient variations of the fine structure constant resulting from the passage of dark matter may be observed in a global network of optical clocks [24]. The detection of the transient effects relies on correlations in the frequency data of worldwide clocks running simultaneously [75]. To obtain a large amount of temporally overlapped data from laboratories worldwide, it is highly desirable that the participating clocks are capable of running continuously for a long period. The continuous operation also makes it possible to search for dark matter interactions on long timescales that could not be studied in previous experiments [75, 76].
In conclusion, we have demonstrated the nearly continuous operation of an Yb optical lattice clock with an uptime of for half a year. This enabled us to link the Yb clock to the SI second with an uncertainty of , and to determine the absolute frequency of the Yb clock transition with a total uncertainty of . The present performance preliminary demonstrated that the robustness of optical lattice clocks can reach a level comparable to that of Cs fountain clocks. This work constitutes an important step towards the redefinition of the SI second.
Acknowledgments
We thank H. Katori, M. Takamoto, and H. Imai for providing information on their vacuum systems, and S. Okubo for technical assistance. We are indebted to national metrology institutes for their efforts in operating the primary and secondary frequency standards and making the data available in Circular T. This work was supported by Japan Society for the Promotion of Science (JSPS) KAKENHI Grant Number 17H01151, and 17K14367, and JST-Mirai Program Grant Number JPMJMI18A1, Japan.
Reference
References
- [1] Bloom B J, Nicholson T L, Williams J R, Campbell S L, Bishof M, Zhang X, Zhang W, Bromley S L, and Ye J 2014 An optical lattice clock with accuracy and stability at the level 71-75
- [2] Ushijima I, Takamoto M, Das M, Ohkubo T, and Katori H 2015 Cryogenic optical lattice clocks 185-189
- [3] Nicholson T L, Campbell S L, Hutson R B, Marti G E, Bloom B J, McNally R L, Zhang W, Barrett M D, Safronova M S, Strouse G F, Tew W L, and Ye J 2015 Systematic evaluation of an atomic clock at total uncertainty 6896
- [4] McGrew W F, Zhang X, Fasano R J, Schäffer S A, Beloy K, Nicolodi D, Brown R C, Hinkley N, Milani G, Schioppo M, Yoon T H, and Ludlow A D 2018 Atomic clock performance enabling geodesy below the centimetre level 87-90
- [5] Bothwell T, Kedar D, Oelker E, Robinson J M, Bromley S L, Tew W L, Ye J, and Kennedy C J 2019 JILA SrI optical lattice clock with uncertainty of 066004.
- [6] Gill P 2011 When should we change the definition of the second? A 4109-4130
- [7] Hong F-L 2016 Optical frequency standards for time and length applications 012002
- [8] Riehle F, Gill P, Arias F, and Robertsson L. 2018 The CIPM list of recommended frequency standard values: Guidelines and procedures 188-200
- [9] Lodewyck J 2019 On a definition of the SI second with a set of optical clock transitions 055009
- [10] Campbell G K, Ludlow A D, Blatt S, Thomsen J W, Martin M J, de Miranda M H G, Zelevinsky T, Boyd M M, Ye J, Diddams S A, Heavner T P, Parker T E, and Jefferts S R 2008 The absolute frequency of the 87Sr optical clock transition 539-548
- [11] Lemke N D, Ludlow A D, Barber Z W, Fortier T M, Diddams S A, Jiang Y, Jefferts S R, Heavner T P, Parker T E, and Oates C W 2009 Spin-1/2 optical lattice clock 063001
- [12] Le Targat R, Lorini L, Le Coq Y, Zawada M, Gueńa J, Abgrall M, Gurov M, Rosenbusch P, Rovera D G, Nagórny B, Gartman R, Westergaard P G, Tobar M E, Lours M, Santarelli G, Clairon A, Bize S, Laurent P, Lemonde P, and Lodewyck J 2013 Experimental realization of an optical second with strontium lattice clocks 2109
- [13] Grebing C, Al-Masoudi A, Dörsher S, Häfner S, Gerginov V, Weyers S, Lipphardt B, Riehle F, Sterr U, and Lisdat C 2016 Realization of a timescale with an accurate optical lattice clock 563-569
- [14] Lodewyck J, Bilicki S, Bookjans E, Robyr J-L, Shi C, Vallet G, Le Targat R, Nicolodi D, Coq Y, Guéna J, Abgrall M, Rosenbusch P, and Bize S 2016 Optical to microwave clock frequency ratios with a nearly continuous strontium optical lattice clock 1123-1130
- [15] Pizzocaro M, Thoumany P, Rauf B, Bregolin F, Milani G, Clivati C, Costanzo G A, Levi F, and Calonico D 2017 Absolute frequency measurement of the 1SP0 transition of 171Yb 102-112
- [16] Panfilo G and Parker T E 2010 A theoretical and experimental analysis of frequency transfer uncertainty, including frequency transfer into TAI 552-560
- [17] Tanabe T, Akamatsu D, Kobayashi T, Takamizawa A, Yanagimachi S, Ikegami T, Suzuyama T, Inaba H, Okubo S, Yasuda M, Hong F-L, Onae A, and Hosaka K 2015 Improved frequency measurement of the 1SP0 clock transition in 87Sr using a Cs fountain clock as a transfer oscillator . . . . 115002
- [18] Hachisu H, Petit G, Nakagawa F, Hanado Y, and Ido T 2017 SI-traceable measurement of an optical frequency at the low level without a local primary standard . 8511-8523
- [19] McGrew W F, Zhang X, Leopardi H, Fasano R J, Nicolodi D, Beloy K, Yao J, Sherman J A, Schäffer S A, Savory J, Brown R C, Römisch S, Oates C W, Parker T E, Fortier T M, and Ludlow A D 2019 Towards the optical second: verifying optical clocks at the SI limit 448-454
- [20] Pizzocaro M, Bregolin F, Barbieri P, Rauf B, Levi F, and Calonico D 2019 Absolute frequency measurement of the 1SP0 transition of 171Yb with a link to International Atomic Time
- [21] Hill I R, Hobson R, Bowden W, Bridge E M, Donnellan S, Curtis E A, and Gill P 2016 A low maintenance Sr optical lattice clock . . . . 012019
- [22] Blatt S, Ludlow A D, Campbell G K, Thomsen J W, Zelevinsky T, Boyd M M, Ye J, Baillard X, Fouché M, Le Targat R, Brusch A, Lemonde P, Takamoto M, Hong F-L, Katori H, Flambaum V V 2008 New Limits on Coupling of Fundamental Constants to Gravity Using 87Sr Optical Lattice Clocks 140801
- [23] Safronova M S, Budker D, DeMille D, Jackson Kimball D F, Derevianko A, and Clark C W 2018 Search for new physics with atoms and molecules 025008
- [24] Derevianko A and Pospelov M 2014 Hunting for topological dark matter with atomic clocks 933-936
- [25] Wcisło P, Morzyński P, Bober M, Cygan A, Lisak D, Ciuryło R, and Zawada M 2016 Experimental constraint on dark matter detection with optical atomic clocks 0009
- [26] Takamoto M, Ushijima I, Ohmae N, Yahagi T, Kokado K, Shinkai H, and Katori H 2020 Test of General Relativity by a Pair of Transportable Optical Lattice Clocks . .
- [27] Takano T, Takamoto M, Ushijima I, Ohmae N, Akatsuka T, Yamaguchi A, Kuroishi, Munekane H, Miyahara B, and Katori H 2016 Geopotential measurements with synchronously linked optical lattice clocks 662-666
- [28] Grotti J . . 2018 Geodesy and metrology with a transportable optical clock 437-441
- [29] Hachisu H, Nakagawa F, Hanado Y, and Ido T 2018 Months-long real-time generation of a time scale based on an optical clock 4243
- [30] Milner W R, Robinson J M, Kennedy C J, Bothwell T, Kedar D, Matei D G, Legero T, Sterr U, Riehle F, Leopardi H, Fortier T M, Sherman J A, Levine J, Yao J, Ye J, and Melker E 2019 Demonstration of a Timescale Based on a Stable Optical Carrier 173201
- [31] Yao J, Sherman J A, Fortier T, Leopardi H, Parker T, McGrew W, Zhang X, Nicolodi D, Fasano R, Schäffer S, Beloy K, Savory J, Romisch S, Oates C, Diddams S, Ludlow A, and Levine J 2019 Optical-Clock-Based Time Scale 044069
- [32] Inaba H, Daimon Y, Hong F-L, Onae A, Minoshima K, Schibli T R, Matsumoto H, Hirano M, Okuno T, Onishi M, and Nakazawa M 2006 Long-term measurement of optical frequencies using a simple, robust and low-noise fiber based frequency comb 5223-5231
- [33] Hisai Y, Akamatsu D, Kobayashi T, Okubo S, Inaba H, Hosaka K, Yasuda M, and Hong F-L 2019 Development of 8-branch Er:fiber frequency comb for Sr and Yb optical lattice clocks 6404-6414
- [34] Kobayashi T, Akamatsu D, Hosaka K, and Yasuda M 2019 A relocking scheme for optical phase locking using a digital circuit with an electrical delay line 103002
- [35] Kobayashi T, Akamatsu D, Hisai Y, Tanabe T, Inaba H, Suzuyama T, Hong F-L, Hosaka K, and Yasuda M 2018 Uncertainty Evaluation of an 171Yb Optical Lattice Clock at NMIJ 2449-2458
- [36] Katori H, Takamoto M, Pal’chikov V G, and Ovsiannikov V D 2003 Ultrastable optical clock with neutral atoms in an engineered light shift trap 173005
- [37] Kobayashi T, Akamatsu D, Hosaka K, Inaba H, Okubo S, Tanabe T, Yasuda M, Onae A, and Hong F-L 2016 Absolute frequency measurements and hyperfine structures of the molecular iodine transitions at 578 nm 725-734
- [38] Saxberg B, Plotkin-Swing B, and Gupta S Active stabilization of a diode laser injection lock 2016 063109
- [39] Kobayashi T, Akamatsu D, Nishida Y, Tatabe T, Yasuda M, Hong F-L, and Hosaka K 2016 Second harmonic generation at 399 nm resonant on the transition of ytterbium using a periodically poled LiNbO3 waveguide 12142-12150
- [40] Iwakuni K, Inaba H, Nakajima Y, Kobayashi T, Hosaka K, Onae A, and Hong F-L 2012 Narrow linewidth comb realized with a mode-locked fiber laser using an intra-cavity waveguide electro-optic modulator for high-speed control 13769-13776
- [41] Inaba H, Hosaka K, Yasuda M, Nakajima Y, Iwakuni K, Akamatsu D, Okubo S, Kohno T, Onae A, and Hong F-L 2013 Spectroscopy of 171Yb in an optical lattice based on laser linewidth transfer using a narrow linewidth frequency comb 7891-7896
- [42] Hisai Y, Ikea K, Sakagami H, Horikiri T, Kobayashi T, Yoshii K, and Hong F-L 2018 Evaluation of laser frequency offset locking using an electrical delay line 5628-5634
- [43] Akamatsu D, Inaba H, Hosaka K, Yasuda M, Onae A, Suzuyama T, Amemiya M, and Hong F-L 2013 Spectroscopy and frequency measurement of the 87Sr clock transition by laser linewidth transfer using an optical frequency comb 012401
- [44] Katori H, Ovsiannikov V D, Marmo S I, and Palchikov V G 2015 Strategies for reducing the light shift in atomic clocks 052503
- [45] Nemitz N, Ohkubo T, Takamoto M, Ushijima I, Das M, Ohmae N, and Katori H 2016 Frequency ratio of Yb and Sr clocks with uncertainty at 150 seconds averaging time 258-261
- [46] Nemitz N, Jørgensen A A, Yanagimoto R, Bregolin F, and Katori H 2019 Modeling light shifts in optical lattice clocks 033424
- [47] Brown R C, Phillips N B, Beloy K, McGrew W F, Schioppo M, Fasano R J, Milani G, Zhang X, Hinkley N, Leopardi H, Yoon T H, Nicolodi D, Fortier T M, and Ludlow A D 2017 Hyperpolarizability and Operational Magic Wavelength in an Optical Lattice Clock 253001
- [48] Blatt S, Thomsen J W, Campbell G K, Ludlow A D, Swallows M D, Martin M J, Boyd M M, and Ye J 2009 Rabi spectroscopy and excitation inhomogeneity in a one-dimensional optical lattice clock 052703
- [49] Kim H, Heo M-S, Lee W-K, Park C Y, Hong H-G, Hwang S-W, and Yu D-H 2017 Improved absolute frequency measurement of the 171Yb optical lattice clock at KRISS relative to the SI second 050302
- [50] Middelmann T, Lisdat C, Falke S, Vellore Winfred J S R, Riehle F, and Sterr U 2011 Tackling the blackbody shift in a strontium optical lattice clock 2550
- [51] Sherman J A, Lemke N D, Hinkley N, Pizzocaro M, Fox R W, Ludlow A D, and Oates C W 2012 High-Accuracy Measurement of Atomic Polarizability in an Optical Lattice Clock 153002
- [52] Beloy K, Hinkley N, Phillips N B, Sherman J A, Shiopppo M, Lehman J, Feldman A, Hanssen L M, Oates C W, Ludlow A D 2014 Atomic Clock with Room-Temperature Blackbody Stark Uncertainty 260801
- [53] Lemke N D, von Stecher J, Sherman J A, Rey A M, Oates C W, and Ludlow A D 2011 Wave Cold Collisions in an Optical Lattice Clock 103902
- [54] Ludlow A D, Lemke N D, Sherman J A, Oates C W, Quéméner G, von Stecher J, and Rey A M 2011 Cold-collision-shift cancellation and inelastic scattering in a Yb optical lattice clock 052724
- [55] Lee S, Park C Y, Lee W-K, and Yu D-H 2016 Cancellation of collisional frequency shifts in optical lattice clocks with Rabi spectroscopy 033030
- [56] Nakajima Y, Inaba H, Hosaka K, Minoshima K, Onae A, Yasuda M, Kohno T, Kawato S, Kobayashi T, Katsuyama T, and Hong F-L 2010 A multi-branch, fiber-based frequency comb with millihertz-level relative linewidths using an intra-cavity electro-optic modulator 1667-1676
- [57] https://www.bipm.org/en/bipm/tai/.
- [58] Yu D-H, Weisse M, and Parker T E 2007 Uncertainty of a frequency comparison with distributed dead time and measurement interval offset 91-96
- [59] Bauch A 2005 The PTB primary clocks CS1 and CS2 S43-S54
- [60] Guéna J, Abgrall M, Rovera D, Laurent P, Chupin B, Lours M, Santarelli G, Rosenbusch P, Tobar M E, Li R, Gibble K, Clairon A, and Bize S 2012 Progress in atomic fountains at LNE-SYRTE 391-409
- [61] Weyers S, Gerginov V, Kazda M, Rahm J, Lipphardt B, Dobrev G, and Gibble K 2018 Advances in the accuracy, stability, and reliability of the PTB primary fountain clocks 789-805
- [62] Blinov I Y, Boiko A I, Domnin Y S, Kostromin V P, Kupalova O V, and Kupalov D S 2017 Budget of Uncertainties in the Cesium Frequency Frame of Fountain Type 30-36
- [63] Fang F, Li M, Lin P, Chen W, Liu N, Lin Y, Wang P, Liu K, Suo R, and Li T 2015 NIM5 Cs fountain clock and its evaluation 454-468
- [64] Guéna J, Abgrall M, Clairon A, and Bize S 2014 Contributing to TAI with a secondary representation of the SI second 108-120
- [65] Hachisu H, Petit G, and Ido T 2016 Absolute frequency measurement with uncertainty below using International Atomic Time 34
- [66] Kohno T, Yasuda M, Hosaka K, Inaba H, Nakajima Y, and Hong F-L 2009 One-Dimensional Optical Lattice Clock with a Fermionic 171Yb Isotope 072501
- [67] Yasuda M, Inaba H, Kohno T, Tanabe T, Nakajima Y, Hosaka K, Akamatsu D, Onae A, Suzuyama T, Amemiya M, and Hong F-L 2012 Improved Absolute Frequency Measurement of the 171Yb Optical Lattice Clock towards a Candidate for the Redefinition of the Second 102401
- [68] Park C Y, Yu D-J, Lee W-K, Park S E, Kim E B, Lee S K, Cho J W, Yoon T H, Mun J, Park S J, Kwon T Y, and Lee S-B 2013 Absolute frequency measurement of 1SP transition of 171Yb atoms in a one-dimensional optical lattice at KRISS 119-128
- [69] Akamatsu D, Yasuda M, Inaba H, Hosaka K, Tanabe T, Onae A, and Hong F-L 2014 Frequency ratio measurement of 171Yb and 87Sr optical lattice clocks 7898-7905
- [70] Takamoto M, Ushijima I, Das M, Nemitz N, Ohkubo T, Yamanaka K, Ohmae N, Takano T, Akatsuka T, Yamaguchi A, and Katori H 2015 Frequency ratios of Sr, Yb, and Hg based optical lattice clocks and their applications 489-498
- [71] Fujieda M, Yang S-H, Gotoh T, Hwang S-W, Hachisu H, Kim H, Lee Y K, Tabuchi R, Ido T, Lee W-K, Heo M-S, Park C Y, Yu D-H, and Petit G 2018 Advanced satellite-based frequency transfer at the level 973-978
- [72] Akamatsu D, Kobayashi T, Hisai Y, Tanabe T, Hosaka K, Yasuda M, and Hong F-L 2018 Dual-mode operation of an optical lattice clock using strontium and ytterbium atoms 1069-1075
- [73] Peil S, Crane S, Hanssen J L, Swanson T B, and Ekstrom C R 2013 Tests of local position invariance using continuously running atomic clocks 010102(R)
- [74] Takamizawa A, Yanagimachi S, Tanabe T, Hagimoto K, Hirano I, Watabe K, Ikegami T, and Hartnett J G 2015 Preliminary Evaluation of the Cesium Fountain Primary Frequency Standard NMIJ-F2 2504-2512
- [75] Wcisło P 2018 New bounds on dark matter coupling from a global network of optical atomic clocks , eaau4869
- [76] Roberts B M 2019 Search for transient variations of the fine structure constant and dark matter using fiber-linked optical atomic clocks arXiv:1907.02661