Demonstration of the nearly continuous operation of an 171Yb optical lattice clock for half a year

Takumi Kobayashi1, Daisuke Akamatsu1, Kazumoto Hosaka1, Yusuke Hisai2, Masato Wada1, Hajime Inaba1, Tomonari Suzuyama1, Feng-Lei Hong2, Masami Yasuda1 1National Metrology Institute of Japan (NMIJ), National Institute of Advanced Industrial Science and Technology (AIST), 1-1-1 Umezono, Tsukuba, Ibaraki 305-8563, Japan
2Department of Physics, Graduate School of Engineering Science, Yokohama National University, 79-5 Tokiwadai, Hodogaya-ku, Yokohama 240-8501, Japan
[email protected]
Abstract

Optical lattice clocks surpass primary Cs microwave clocks in frequency stability and accuracy, and are promising candidates for a redefinition of the second in the International System of Units (SI). However, the robustness of optical lattice clocks has not yet reached a level comparable to that of Cs fountain clocks which contribute to International Atomic Time (TAI) by the nearly continuous operation. In this paper, we report the long-term operation of an 171Yb optical lattice clock with a coverage of 80.3%80.3\% for half a year including uptimes of 93.9%93.9\% for the first 24 days and 92.6%92.6\% for the last 35 days. This enables a nearly dead-time-free frequency comparison of the optical lattice clock with TAI over months, which provides a link to the SI second with an uncertainty of low 101610^{-16}. By using this link, the absolute frequency of the 1S30{}_{0}-^{3}P0 clock transition of 171Yb is measured as 518 295 836 590 863.54(26)518\,295\,836\,590\,863.54(26) Hz with a fractional uncertainty of 5.0×10165.0\times 10^{-16}. This value is in agreement with the recommended frequency of 171Yb as a secondary representation of the second.

1 Introduction

Several optical lattice clocks have achieved fractional frequency stabilities and uncertainties at the 101810^{-18} level [1, 2, 3, 4, 5], which is better than primary Cs fountain microwave clocks. These achievements have stimulated discussion regarding a redefinition of the second in the International System of Units (SI) [6, 7, 8, 9].

With the aim of redefining the SI second, the absolute frequencies of optical lattice clocks have been measured by many groups and used to determine the recommended frequencies of secondary representations of the second [8]. The absolute frequency can be directly measured referenced to a Cs fountain clock if it is locally available in the laboratory [14, 11, 10, 12, 13, 15]. When a local Cs clock is not available, the absolute frequency can be measured using a satellite link to International Atomic Time (TAI). TAI is a global timescale computed by the Bureau International des Poids et Mesures (BIPM). BIPM provides the frequency difference between TAI and the SI second averaged over a month. Therefore, the continuous comparison of an optical clock with TAI for a month is desirable in terms of linking to the SI second without additional uncertainty resulting from the dead time. A long-term comparison is also needed to reduce the uncertainty arising from the satellite link [16]. Since optical lattice clocks have commonly been operated intermittently, previous works have mostly employed stable local flywheels (e.g, an ensemble of hydrogen masers) that reach the 101610^{-16} level to bridge the gaps in the operation of the lattice clocks [17, 18, 19, 20]. A few groups have reported the nearly continuous operation of Sr optical lattice clocks with uptimes of 93%93\% for 10 days, 83%83\% for three weeks [14], and 84%84\% for 25 days [21].

The continuous operation of an optical lattice clock is essential for the calibration of the frequency of TAI as a secondary representation of the second [8] as well as the absolute frequency measurement. A robust lattice clock is also important for new applications including tests of fundamental physics [12, 19, 22, 23, 24, 25, 26], relativistic geodesy [27, 28], and the generation of a stable local timescale [13, 29, 30, 31].

The main challenge as regards realizing the continuous operation of an optical lattice clock is to stabilize the frequencies of several light sources including an optical frequency comb to allow the comparison of the optical clock and a microwave standard. So far, we have developed multi-branch erbium-doped-fiber-based frequency combs that can run continuously for long periods [32]. By stabilizing the frequencies of many light sources to the comb, we have constructed a simple and reliable laser system for clock experiments [33]. We have also developed a relocking scheme [34] for optical phase locking. These techniques should greatly assist the realization of the continuous operation.

In this paper, we report the operation of an 171Yb optical lattice clock with a coverage of 80.3%80.3\% for half a year (185 days) including uptimes of 93.9%93.9\% for the first 24 days, 86.4%86.4\% for the second 27 days, 80.4%80.4\% for the third 30 days, 72.7%72.7\% for the fourth 35 days, 82.6%82.6\% for the fifth 25 days, and 92.6%92.6\% for the sixth 35 days. Using a single hydrogen maser with its stability limited by a flicker floor of 2×10152\times 10^{-15}, we reduce the uncertainty in the link between the Yb clock and the SI second to the low 101610^{-16} level, and measure the absolute frequency of the Yb clock transition.

2 Experimental setup

Refer to caption
Figure 1: Experimental setup. PDH: Pound-Drever-Hall stabilization, PL: Phase lock, ECDL: External cavity diode laser, PPLN: periodically poled LiNbO3, Ti:S: Titanium sapphire, SOA: Semiconductor optical amplifier, FL: Frequency lock, BD: Beat detection, FD: Fluorescence detection.

Figure 1 is a schematic diagram of the experimental setup. The setup for the Yb optical lattice clock is similar to that reported in our previous paper [35]. The clock utilizes the 1S30{}_{0}-^{3}P0 clock transition at 578 nm of 171Yb atoms trapped in an optical lattice operated at the magic wavelength [36] of 759 nm. The 578 nm light was generated by the second harmonic generation (SHG) of an external cavity diode laser (ECDL) emitting at 1156 nm [37]. The lattice light at 759 nm was provided by a titanium sapphire (Ti:S) laser (M squared, SolsTiS-4000-SRX-R). Before loading the atoms to the optical lattice, two-stage laser cooling with the 1S10{}_{0}-^{1}P1 transition at 399 nm and the 1S30{}_{0}-^{3}P1 transition at 556 nm was carried out to decrease the temperature of the atoms to several tens of microkelvins. The 399 nm light was generated by an injection-locked 399 nm diode laser with an automatic relocking mechanism [38]. A seed light for the injection locking was provided by the SHG of a 798 nm ECDL [39]. The 556 nm light for the second stage cooling was prepared by the SHG of a 1112 nm light from a fiber laser (NKT Photonics, Koheras) with a semiconductor optical amplifier. All the SHG processes were performed with single-pass periodically poled LiNbO3 waveguides.

The 1112 nm fiber laser and the 1156 nm ECDL were phase locked to a narrow linewidth fiber comb (Fiber comb 1 in Fig. 1) [41, 40]. This comb was phase locked to a master Nd:YAG laser at 1064 nm, which was stabilized to an ultra-low expansion (ULE) cavity by using the Pound-Drever-Hall method. To compensate for the narrow capture range of optical phase locking, a recently-developed relocking scheme [34] was incorporated in the phase locking of the comb to the master laser and the phase locking of the 1112 nm and 1156 nm lasers to the comb. The 798 nm ECDL and the Ti:S laser were stabilized to another fiber comb (Fiber comb 2 in Fig. 1) by frequency locking with an electrical delay line [42], which ensured a large capture range. Fiber comb 2 was referenced to the Coordinated Universal Time of the National Metrology Institute of Japan (UTC(NMIJ)), which is generated by a hydrogen maser and an auxiliary output generator to add an arbitrary frequency offset.

The Yb optical lattice clock was operated with a cycle time of about 1.7 s. Yb atoms effused from an oven were decelerated with a Zeeman slower and cooled in first-stage cooling for about 1.2 s and then further cooled in second-stage cooling for 275 ms. In the present work, a window heated at 204C was installed inside a vacuum chamber in the pass of the Zeeman slower beam to prevent the atomic beam from coating the viewport. The atoms were then loaded into a vertically oriented one-dimensional optical lattice and spin-polarized with the 1S30{}_{0}-^{3}P1 transition at 556 nm for 20 ms. A Rabi π\pi pulse resonant on the clock transition was then applied for 40 ms. The atomic population and the excitation probability were deduced using a laser-induced fluorescence signal by the 1S10{}_{0}-^{1}P1 transition at 399 nm, and repumping on the 3P30{}_{0}-^{3}D1 transition at 1389 nm induced by a free-running distributed feedback laser. The excitation probability was used to calculate frequency corrections applied for an acousto-optic modulator (AOM), which steers the frequency of the clock laser to the atomic transition. The atomic transition was split into two Zeeman components mF=±1/2m_{F}=\pm 1/2 by applying a bias magnetic field of 65 μ\muT. The clock laser was stabilized to each component.

Fiber comb 2 was also used to measure the frequency of the Yb clock against UTC(NMIJ). To transfer the stability of UTC(NMIJ) to the comb without degradation, we mixed the 7th harmonics of the repetition rate frequency frep121.8f_{\mathrm{rep}}\sim 121.8 MHz with a 900 MHz signal which was generated by the 90-fold frequency multiplication of a 10 MHz signal from UTC(NMIJ). The resulting output frequency from the mixer δ47\delta\sim 47 MHz was phase locked to a synthesizer referenced to UTC(NMIJ). After stabilizing the comb to UTC(NMIJ), the beat frequency between the clock laser and the comb was measured by using a zero dead-time frequency counter.

3 Demonstration of the nearly continuous operation

The frequency of the Yb optical lattice clock was measured against UTC(NMIJ) during a half-year period from the Modified Julian Date (MJD) 58754 (28 September 2019) to MJD 58939 (31 March 2020). The Yb clock was operated with uptimes of 93.9%\% in the first 24-day campaign from MJD 58754 to MJD 58778, 86.4%\% in the second 27-day campaign from MJD 58787 to MJD 58814, 80.4%\% in the third 30-day campaign from MJD 58814 to MJD 58844, 72.7%\% in the fourth 35-day campaign from MJD 58844 to MJD 58879, 82.6%\% in the fifth 25-day campaign from MJD 58879 to MJD 58904, and 92.6%\% in the sixth 35-day campaign from MJD 58904 to MJD 58939 (see also Table 1). Figures 2 (a) - (f) respectively show the fractional frequency differences between the Yb clock and UTC(NMIJ), denoted by y(YbUTC(NMIJ))y(\mathrm{Yb}-\mathrm{UTC(NMIJ)}), in the first - sixth campaigns. The uptime was calculated by using NvalidTcycle/TtotalN_{\mathrm{valid}}T_{\mathrm{cycle}}/T_{\mathrm{total}}, where NvalidN_{\mathrm{valid}} is the number of valid data points, TcycleT_{\mathrm{cycle}} the clock cycle time (1.7\sim 1.7 s), and TtotalT_{\mathrm{total}} the total period of the measurement. We discarded data that included the following events: (i) large excursions in the phase-locked frequencies that can occur during the relocking procedures; (ii) cycle slips in the frequency counting; and (iii) low excitation probabilities of the clock transition that lasted for a sufficiently long period.

The fractional frequency difference y(YbUTC(NMIJ))y(\mathrm{Yb}-\mathrm{UTC(NMIJ)}) is given by

y(YbUTC(NMIJ))\displaystyle y(\mathrm{Yb}-\mathrm{UTC(NMIJ)}) =\displaystyle= fa(Yb)fn(Yb)fa(UTC(NMIJ))fn(UTC(NMIJ))\displaystyle\frac{f^{\mathrm{a}}(\mathrm{Yb})}{f^{\mathrm{n}}(\mathrm{Yb)}}-\frac{f^{\mathrm{a}}(\mathrm{UTC(NMIJ))}}{f^{\mathrm{n}}(\mathrm{UTC(NMIJ)})} (1)
\displaystyle\approx fa(Yb)/fa(UTC(NMIJ))fn(Yb)/fn(UTC(NMIJ))1,\displaystyle\frac{f^{\mathrm{a}}(\mathrm{Yb})/f^{\mathrm{a}}(\mathrm{UTC(NMIJ))}}{f^{\mathrm{n}}(\mathrm{Yb)}/f^{\mathrm{n}}(\mathrm{UTC(NMIJ)})}-1,\quad

where fa(n)(X)f^{\mathrm{a(n)}}(\mathrm{X}) denotes the actual (nominal) frequency of X. The approximation is vaild when (fa(X)fn(X))fn(X)(f^{\mathrm{a}}(X)-f^{\mathrm{n}}(X))\ll f^{\mathrm{n}}(X). Here we chose fn(UTC(NMIJ))=10f^{\mathrm{n}}(\mathrm{UTC(NMIJ)})=10 MHz and fn(Yb)=fCIPM(Yb)=518 295 836 590 863.6f^{\mathrm{n}}(\mathrm{Yb)}=f^{\mathrm{CIPM}}(\mathrm{Yb})=518\,295\,836\,590\,863.6 Hz, which is the CIPM (Comité International des Poids et Mesures) recommended frequency of 171Yb [8].

Table 1: Uptimes of the Yb optical lattice clock
Campaign MJD Period (days) Uptime (%)(\%)
1 58754 - 58778 24 93.993.9
2 58787 - 58814 27 86.486.4
3 58814 - 58844 30 80.480.4
4 58844 - 58879 35 72.772.7
5 58879 - 58904 25 82.682.6
6 58904 - 58939 35 92.692.6
Total 58754 - 58939 185 80.380.3
Refer to caption
Figure 2: Fractional frequency of the Yb clock relative to UTC(NMIJ) (y(YbUTC(NMIJ)))(y(\mathrm{Yb}-\mathrm{UTC(NMIJ)})) and the uptime per day in the (a) first, (b) second, (c) third, (d) fourth, (e) fifth, and (f) sixth campaigns. The blue point corresponds to a 6.8 s average, the red point a 10310^{3} s average, and the gray point in (f) a 30 s average. TlinkT_{\mathrm{link}} indicates the period employed to calculate the frequency difference between UTC(NMIJ) and TAI. The data in the shaded region in (e) were not used for the calculation due to a large excursion of the phase of UTC(NMIJ) at MJD 58896.

The clock was mostly unattended during the full campaign period. Several experimental parameters were monitored remotely. A data acquisition computer automatically sent an email alert to operators when one of the experimental parameters was outside the normal range. In the first campaign (see Fig. 2 (a)), to demonstrate almost dead-time-free operation ( 90%\geq 90\% uptime per day), we always carried out manual tuning as soon as we received the email alert. Significant reductions in the uptime were caused by rare events such as the passage of a typhoon at MJD 58770 and the breaking of a diode laser used in fiber comb 1 at MJD 58778. In the second to sixth campaigns (see Figs. 2 (b)-(f)), manual recovery was not always performed at night to demonstrate that a large uptime can also be achieved with minimum human effort. Large gaps from MJD 58831 to MJD 58834 in the third campaign were caused by the shutdown of the air conditioners for facility maintenance. In the fourth campaign, a diode laser used in fiber comb 2 was broken, which caused a large downtime from MJD 58873 to MJD 58875. At MJD 58877, we stopped the clock operation and carried out a systematic evaluation of the microwave synthesis (See Section 5).

Figure 3 shows the distribution of interrupting events that occurred during the full campaign period. The number of interrupting events were counted according to the email alert history. We could basically link the email history to the types of interruptions, but could not link some events due to a lack of records. These events are indicated by “No record” in Fig. 3. The clock operation was mostly interrupted by acoustic noise from a drainage pump in the facility (26%26\% of the total number of interruptions), the instability of the injection locking (12%12\%), mode hopping of the Ti:S laser (11%11\%) and the 1156 nm ECDL (9%9\%), and a small number of trapped atoms (8%8\%). Earthquakes (6%6\%) stopped the Yb clock mostly by dropping the lock of the 798 nm diode laser without implementing the relocking scheme. The other relockable light sources were resistant to small magnitude earthquakes. The other factors that interrupted the operation included rapid frequency drifts of the ULE cavity after the air conditioners were shutdown (MJD 58831 to MJD 58834) and temporary freezing of the data acquisition computer.

Refer to caption
Figure 3: Distribution of interrupting events during the full campaign period.
Refer to caption
Figure 4: The number of interrupting events in each campaign.

Figure 4 shows the number of interrupting events in each campaign. Most of the interruptions occurred with a small number (10\lesssim 10) in each campaign except for the Ti:S laser (blue square), a small number of trapped atoms (green square), and a drainage pump (red square). The lock of the Ti:S laser failed frequently especially in the second campaign. This problem was partially solved in the following campaigns by cleaning the mirrors used to introduce the pump laser beam and improving the signal-to-noise ratio of the error signal for the frequency locking to the comb. From the second campaign, the number of atoms trapped in the optical lattice gradually decreased (see Section 4), which increased the occurrence of lock failures. From the end of the third campaign, the Yb clock was frequently interrupted by a drainage pump, which had been installed in the facility during the third campaign. The pump generated acoustic noise for about 30 minutes once a day when it drained water. During this period, the excitation probability of the atomic transition became almost zero due to the broadening of the clock laser linewidth.

4 Frequency stability and uncertainty of the optical lattice clock

Refer to caption
Figure 5: Allan deviation of UTC(NMIJ) referenced to the Yb clock (red dots) and the frequency ratio Yb/Sr (blue triangles). The green curve shows the stability of a simulated model of UTC(NMIJ) used to calculate the dead time uncertainty (see text). The purple line indicates a slope of 1.0×1014/(τ/s)1.0\times 10^{-14}/\sqrt{(\tau/\mathrm{s})}.
Table 2: Systematic shifts and uncertainties of the 171Yb optical lattice clock obtained in the first campaign from MJD 58754 to MJD 58778.
Effect Shift Uncertainty
(×1017\times 10^{-17}) (×1017\times 10^{-17})
Lattice light 3.4 33.1
BBR 263.8-263.8 20.8
Density 8.3-8.3 6.4
Second-order Zeeman 5.2-5.2 0.3
Probe light 0.4 0.2
Servo error 4.7-4.7 1.1
AOM switching - 1
Total 278.3-278.3 39.6
Refer to caption
Figure 6: Atomic density (ρ1014\rho\sim 10^{14} m-3) in the optical lattice (blue points), Zeeman splitting (pink points), excitation probability (green points) in the (a) first, (b) second, (c) third, (d) fourth, (e) fifth, and (f) sixth campaigns. The data point corresponds to a 6.8 s average. Some of the data at MJD 58789 in (b) were not recorded. The data in the shaded region (MJD 58894 to MJD 58899 in (e)) were not used for the systematic evaluation. The fiber for the 399 nm light was replaced at MJD 58794 in (b), MJD 58814 in (c), MJD 58879 in (e), and MJD 58917 in (f). The 1112 nm semiconductor amplifier was replaced at MJD 58844 in (d), MJD 58853 in (d), MJD 58891 in (e), and MJD 58913 in (f).

We evaluated the frequency stability of the Yb optical lattice clock by comparing it with our Sr optical lattice clock [43]. Figure 5 shows the Allan deviation of the frequency ratio Yb/Sr, which was improved with a slope of 1.0×1014/(τ/s)1.0\times 10^{-14}/\sqrt{(\tau/\mathrm{s})} and reached low 101610^{-16} in several thousand seconds. Figure 5 also shows the frequency stability of UTC(NMIJ) relative to the Yb clock, which was calculated from the data of the first campaign (see Fig. 2 (a)). During this campaign period, the frequency offset of the auxiliary output generator for UTC(NMIJ) was constant.

Table 2 lists the systematic shifts and uncertainties of the Yb optical lattice clock obtained in the first campaign. The uncertainty evaluation is mostly based on our previous evaluation [35]. The density shift and the servo error differed slightly depending on the data obtained in each campaign.

The lattice induced light shift was estimated using a model [44, 45, 46] that includes the contributions from the dominant electric-dipole (E1E1) polarizability, the multipolar (M1M1 and E2E2) polarizability, and hyperpolarizability [47]. To calculate the shift using this model, the effective trap potential depth VeV_{\mathrm{e}}, which is less than the maximum trap depth V0=450ErV_{0}=450E_{r} due to the radial motion of the atoms, and the average vibrational quantum number n\Braket{n} were estimated by sideband spectroscopy of the clock transition [48]. The obtained values were Ve=342(28)ErV_{\mathrm{e}}=342(28)E_{r} and n=2.1(6)\Braket{n}=2.1(6). The frequency of the lattice laser was stabilized at νl=394 798 263\nu_{\mathrm{l}}=394\,798\,263 MHz.

To find the E1E1 polarizability coefficient aa and the E1E1 magic frequency νE1\nu_{\mathrm{E1}}, we have previously measured the light shift as a function of the frequency of the lattice laser [35], and fitted the obtained data with the light shift model [45]. To estimate the variation of aa and νE\nu_{\mathrm{E}} resulting from the systematic uncertainty of the fixed parameters in the light shift model (VeV_{e}, n\Braket{n}, and the multipolar and hyperpolarizability shifts), we have employed a Monte Carlo method in which the fittings were repeated with different permutations of the fixed parameters. Since the distribution of aa obtained from many fittings was asymmetric and had a very long tail mainly due to a relatively large uncertainty in our estimation of VeV_{\mathrm{e}}, we have previously calculated a mean value of aa by removing rare events in the tail. Here, we employed a value of aa that was determined with center values of the fixed parameters. The uncertainty of aa was estimated from a 68%68\% coverage interval of the distribution of aa. The same method was applied to the estimation of νE1\nu_{\mathrm{E1}}. We obtained a=0.020(13)a=0.020(13) mHz/(MHzErE_{r}) and νE1=394 798 244(25)\nu_{E1}=394\,798\,244(25) MHz, which agreed with our previous estimation [35] and the values reported by other groups [46, 47, 49, 20]. Finally, the light shift was estimated to be 3.4(33.1)×10173.4(33.1)\times 10^{-17} with aa, νE1\nu_{E1}, νl\nu_{\mathrm{l}}, n\Braket{n}, VeV_{\mathrm{e}} estimated here, and the sensitively coefficients for the multipolar and hyperpolarizability shifts used in our previous evaluation [35].

The blackbody radiation (BBR) shift was estimated to be 263.8(20.8)×1017-263.8(20.8)\times 10^{-17}, which includes the contribution from the vacuum chamber and the atomic oven considered in our previous evaluation [35] and also that from the newly-installed heated window at 204C. To include this new contribution, the BBR shift induced by the heated window was calculated using a model [50] in which the atoms are located at the center of a stainless-steel sphere with an emissivity of 0.1. The radius of the sphere is equal to the distance between the atoms and the heated window (140\sim 140 mm). The BBR photons at 204C are provided from a small portion of the sphere’s surface with its area equal to that of the heated window seen by the atoms (260\sim 260 mm2). This model takes into account the diffuse reflection of the BBR photons on the stainless-steel wall, which increases the effective solid angle by a factor of 10\sim 10 compared with the geometric solid angle. With the sensitivity coefficient [51, 52], the calculated shift due to the heated window was 1.8×1016-1.8\times 10^{-16}. This shift value was also conservatively taken as an uncertainty.

The density shift was estimated from the atomic density in each campaign. The blue points in Fig. 6 show the atomic density data. The density decreased with a time scale of 10\sim 10 days largely due to a reduction in the transmission efficiency of the 399 nm light through the optical fiber and the output power of the 1112 nm semiconductor amplifier. Some of these components were replaced (see the caption of Fig. 6) when we could not keep the stabilization of the clock laser to the atomic transition. The status of the stabilization was confirmed by monitoring the Zeeman splittings and the excitation probabilities (see pink and green points in Fig. 6). The short-term (a few days) fluctuation of the density was mainly caused by variations in the spectral purity of the injection locking, the polarization rotation of the cooling lasers through the optical fiber, and pointing instabilities of the cooling beam at 399 nm induced by an AOM or a damaged optical fiber. In the first campaign (see blue points in Fig. 6 (a)), the mean value of the atomic density was 3.6(1.6)ρ3.6(1.6)\rho with ρ1014\rho\sim 10^{14} m-3. The uncertainty of 1.6ρ1.6\rho was estimated from the standard deviation. We also added an uncertainty of 10%\sim 10\% to the estimation of the density according to the fluctuation of the 399 nm probe power. With our sensitivity coefficient [35] and a scaling of the trap volume with V03/2V_{0}^{-3/2} [3], the shifts were determined as 8.3(6.4)×1017-8.3(6.4)\times 10^{-17}, 4.9(3.4)×1017-4.9(3.4)\times 10^{-17}, 5.7(4.1)×1017-5.7(4.1)\times 10^{-17}, 3.9(3.0)×1017-3.9(3.0)\times 10^{-17}, 4.5(3.5)×1017-4.5(3.5)\times 10^{-17}, and 5.3(3.7)×1017-5.3(3.7)\times 10^{-17} in the first, second, third, fourth, fifth, and sixth campaigns, respectively. We observed that the mean excitation probability changed from 0.42 to 0.50 after the air conditioners were shutdown (MJD 58831 to MJD 58834) (see Fig. 6 (c)). We attributed this to a variation of the spin-polarization efficiency due to a large frequency change of the ULE cavity. This can change the shift coefficient, since the density shift depends on (i) the excitation probability mostly due to the pp-wave interaction [53, 54, 55] and (ii) the impurity of the spin-polarization [54, 4]. Based on the measured results of Ref. [54], we estimated that in our relatively low atomic density, the change due to those effects is negligibly small compared with the uncertainty in our evaluation.

The second-order Zeeman shift was estimated from the Zeeman splitting measured in each campaign (see pink points in Fig. 6). In the first campaign, the mean value of the splitting was 268(3) Hz, where the uncertainty was estimated from the standard deviation. The shift was found to be 5.2(3)×1017-5.2(3)\times 10^{-17} using our sensitivity coefficient [35]. The mean splitting value was almost constant (268 - 270 Hz) during the entire campaign period.

The probe light shift was calculated to be 4(2)×10184(2)\times 10^{-18} from the shift value of Ref. [4] and the fact that the laser intensity adjusted for the π\pi pulse was inversely proportional to the time duration of the pulse.

The servo error was estimated by averaging the differences between the excitation probabilities of the high- and low-spectral shoulders for the data obtained in each measurement campaign. The servo errors were estimated to be 4.7(1.1)×1017-4.7(1.1)\times 10^{-17}, 4.8(1.0)×1017-4.8(1.0)\times 10^{-17}, 4.6(10.8)×10174.6(10.8)\times 10^{-17}, 2.8(1.2)×1017-2.8(1.2)\times 10^{-17}, 6.0(0.9)×1017-6.0(0.9)\times 10^{-17}, and 4.1(1.4)×1017-4.1(1.4)\times 10^{-17} in the first, second, third, fourth, fifth, and sixth campaigns, respectively. The relatively large uncertainty of 10.8×101710.8\times 10^{-17} in the third campaign arose from the rapid frequency drifts of the ULE cavity after the air conditioners were shutdown (MJD 58831 to MJD 58834).

5 Uncertainty of the frequency comparison by the comb

Refer to caption
Figure 7: Allan deviation calculated from the frequency difference between two independent combs except for the referece signal of UTC(NMIJ). The blue curve indicates the fit of a function which includes the white phase, the white frequency, and the flicker frequency noise components.

To evaluate the uncertainty of the comb-based frequency comparison between the Yb clock and UTC(NMIJ), we compared fiber comb 2 (see Fig. 1) with another comb (fiber comb 3, not shown in Fig. 1) that was independent except for the 10 MHz reference signal of UTC(NMIJ). These combs had independent setups for the microwave synthesis including the 90-fold frequency multiplication. We simultaneously measured the frequency of an acetylene-stabilized laser at 1.5 μ\mum against UTC(NMIJ) by using these two combs, and calculated the frequency difference between the combs. In the first measurement from MJD 58712 to MJD 58713, the fractional frequency difference (fiber comb 2 - fiber comb 3) was found to be 9.7×10179.7\times 10^{-17}, which was the average of the frequency difference data obtained for 105\sim 10^{5} s. Figure 7 shows the Allan deviation calculated from the frequency difference data. The Allan deviation is likely limited by a flicker floor of low 101610^{-16} at averaging times τ104\tau\geq 10^{4} s. By fitting the Allan deviation with a function which includes the white phase, the white frequency, and the flicker frequency noise components, the flicker component was estimated to be 1.6×10161.6\times 10^{-16}. In the second measurement at MJD 58877, the frequency difference and the flicker component were 1.1×1016-1.1\times 10^{-16} and 4.5×10164.5\times 10^{-16}, respectively. We took an average of the two flicker noise values (3.1×10163.1\times 10^{-16}) and divided it by 2\sqrt{2}. The obtained value 2.2×10162.2\times 10^{-16} was adopted as an uncertainty of the frequency comparison. Previous optical-optical comparisons [56] have shown that the uncertainty resulting from the comb itself is negligibly small. We also carried out another measurement in which the two combs shared the common frequency multiplier, and observed that the Allan deviation was improved with a τ1\tau^{-1} slope for τ>103\tau>10^{3} s. Thus, we attributed the flicker noise mainly to the frequency multiplier.

6 Absolute frequency measurement

Since UTC(NMIJ) was continuously compared with TAI via a satellite link (see Fig. 1), the frequency of the Yb clock referenced to the SI second y(YbSI)y(\mathrm{Yb}-\mathrm{SI}) can be deduced from the data y(YbUTC(NMIJ))y(\mathrm{Yb}-\mathrm{UTC(NMIJ)}) in Fig. 2 by the relationship,

y(YbSI)\displaystyle y(\mathrm{Yb}-\mathrm{SI}) =\displaystyle= y(YbUTC(NMIJ))\displaystyle y(\mathrm{Yb}-\mathrm{UTC(NMIJ)}) (2)
+y(UTC(NMIJ)TAI)\displaystyle+y(\mathrm{UTC(NMIJ)}-\mathrm{TAI})
+y(TAISI),\displaystyle+y(\mathrm{TAI}-\mathrm{SI}),

where y(UTC(NMIJ)TAI)y(\mathrm{UTC(NMIJ)}-\mathrm{TAI}) and y(TAISI)y(\mathrm{TAI}-\mathrm{SI}) denote the fractional frequency differences between (i) UTC(NMIJ) and TAI and (ii) TAI and SI, respectively. These values are provided in Circular T [57], which is a monthly report issued by BIPM. Since BIPM only computes y(UTC(NMIJ)TAI)y(\mathrm{UTC(NMIJ)}-\mathrm{TAI}) for 5-day intervals, we chose a 25-day period (MJD 58754 - MJD 58779), a 30-day period (MJD 58784 - MJD 58814), a 30-day period (MJD 58814 - MJD 58844), a 35-day period (MJD 58844 - MJD 58879), a 20-day period (MJD 58879 - MJD 58884 and MJD 58899 - MJD 58904), and a 35-day period (MJD 58904 - MJD 58939) for the first - sixth campaigns, respectively, to calculate the average value of y(UTC(NMIJ)TAI)y(\mathrm{UTC(NMIJ)}-\mathrm{TAI}). These periods are indicated by TlinkT_{\mathrm{link}} in Fig. 2. In the fifth campaign, a 5-day period from MJD 58894 to MJD 58899 (see the shaded region of Fig. 2 (e)) was excluded, since a large excursion of the phase of UTC(NMIJ) occurred during maintenance of the hydrogen maser at MJD 58896. Figure 8 shows the y(YbSI)y(\mathrm{Yb}-\mathrm{SI}) value obtained for each campaign.

Refer to caption
Figure 8: Fractional frequency of the Yb clock referenced to the SI second (y(YbSI))(y(\mathrm{Yb}-\mathrm{SI})) obtained in each campaign. The error bar is given by the total statistical uncertainty for each campaign. The solid red line indicates the weighted mean, the dashed blue lines indicate the total statistical uncertainty in the full period, and the dashed red lines indicate the total uncertainty including the systematic uncertainties.
Table 3: Uncertainty budget for the absolute frequency measurement of the 171Yb clock transition.
Effect First campaign Full period
(×1016\times 10^{-16}) (×1016\times 10^{-16})
Statistics
Yb 0.070.07 0.030.03
Dead time in Yb - UTC(NMIJ) 2.0 0.6
Dead time in TAI - SI 1.6 0.4
UTC(NMJI) - TAI satellite link 2.3 0.8
TAI - SI 0.9 0.4
Statistics total 3.6 1.8
Systematics
Yb 4.04.0 4.04.0
Microwave synthesis 2.2 2.2
TAI - SI 1.1 1.1
Gravitational red shift 0.6 0.6
Systematics total 4.7 4.7
Total 5.9 5.0

The choice of the period TlinkT_{\mathrm{link}} causes an uncertainty due to the dead time in the comparison of the Yb clock and UTC(NMIJ). This dead time uncertainty was calculated by using a numerical simulation described in Ref. [58]. The simulation generated time series data of the frequency according to the noise characteristics: 1×1012/(τ/s)1\times 10^{-12}/(\tau/\mathrm{s}) for the white phase modulation, 7×1014/(τ/s)7\times 10^{-14}/\sqrt{(\tau/\mathrm{s})} for the white frequency modulation (FM), 2×10152\times 10^{-15} for the flicker FM, and 4×1024(τ/s)4\times 10^{-24}\sqrt{(\tau/\mathrm{s})} for the random walk FM. These noise parameters were chosen to reproduce the typical stability of UTC(NMIJ) (see green curve in Fig. 5). We generated two hundred time series patterns and calculated the standard deviation between the frequency averaged over the entire 25-, 30-, or 35-day period and the mean frequency for the time during which the Yb clock was operated. This standard deviation corresponds to the dead time uncertainty. The dead time uncertainties were estimated as 2.0×10162.0\times 10^{-16}, 3.0×10163.0\times 10^{-16}, 3.3×10163.3\times 10^{-16}, 3.3×10163.3\times 10^{-16}, 1.1×10161.1\times 10^{-16}, and 8×10178\times 10^{-17} for the first, second, third, fourth, fifth, and sixth campaigns, respectively. A frequency correction arose from the dead time when the frequency steering of UTC(NMIJ) was carried out. We added corrections of 9.9(2)×1018-9.9(2)\times 10^{-18}, 1.0(9)×1016-1.0(9)\times 10^{-16}, and 7.9(5)×10177.9(5)\times 10^{-17} in the third, fifth, and sixth campaigns, respectively.

In addition, the uncertainty resulting from the dead time when we compare TAI and SI arises from the fact that BIPM calculates the y(TAISI)y(\mathrm{TAI}-\mathrm{SI}) value based on an evaluation with primary and secondary frequency standards for 25, 30, or 35 days, whereas our measurement period TlinkT_{\mathrm{link}} did not fully cover the evaluation period of BIPM in the first and fifth campaigns. This uncertainty was estimated to be 1.6×10161.6\times 10^{-16} in the first campaign, and 4×10174\times 10^{-17} in the full period based on the stability of the Echelle Atomique Libre (EAL) given in Circular T: 1×1015/(τ/day)1\times 10^{-15}/\sqrt{(\tau/\mathrm{day})} for the white FM, 0.35×10150.35\times 10^{-15} for the flicker FM, and 0.2×1016(τ/day)0.2\times 10^{-16}\sqrt{(\tau/\mathrm{day})} for the random walk FM.

The satellite link uncertainty in y(UTC(NMIJ)TAI)y(\mathrm{UTC(NMIJ)}-\mathrm{TAI}) was calculated to be 2.3×10152.3\times 10^{-15} for Tlink=25T_{\mathrm{link}}=25 days, 2.0×10162.0\times 10^{-16} for Tlink=30T_{\mathrm{link}}=30 days, and 1.7×10161.7\times 10^{-16} for Tlink=35T_{\mathrm{link}}=35 days using the recommended formula (2uA)/[(3600×24×5)(Tlink5)0.9](\sqrt{2}u_{\mathrm{A}})/[(3600\times 24\times 5)(\frac{T_{\mathrm{link}}}{5})^{0.9}] [16] with the statistical uncertainty uA=0.3u_{\mathrm{A}}=0.3 ns reported in Circular T.

The statistical and systematic uncertainties of y(TAISI)y(\mathrm{TAI}-\mathrm{SI}) were calculated from the total uncertainty of y(TAISI)y(\mathrm{TAI}-\mathrm{SI}) and the systematic uncertainties of the primary and secondary frequency standards reported in Circular T. During the present half-year period, the frequency of TAI was calibrated by using the following frequency standards: Cs thermal beam clocks (PTB-Cs1 and PTB-Cs2 [59]), Cs fountain clocks (SYRTE-FO1, SYRTE-FO2, SYRTE-FOM [60], PTB-CSF1, PTB-CSF2 [61], SU-CsFO2 [62], and NIM5 [63]), a Rb fountain clock (SYRTE-FORb [64]), and a Sr optical lattice clock (NICT-Sr1 [29]). Since Circular T only provides the total uncertainty of y(TAISI)y(\mathrm{TAI}-\mathrm{SI}), the systematic uncertainty of y(TAISI)y(\mathrm{TAI}-\mathrm{SI}) was estimated using the systematic uncertainties and weights of the individual frequency standards based on a method described in Ref. [65]. The estimated systematic uncertainty of y(TAISI)y(\mathrm{TAI}-\mathrm{SI}) changed between 1.1×10161.1\times 10^{-16} and 1.2×10161.2\times 10^{-16} during the half year period. We adopted an average value of 1.1×10161.1\times 10^{-16} for the full campaign period. The statistical uncertainty of y(TAISI)y(\mathrm{TAI}-\mathrm{SI}) was calculated with the above systematic uncertainty to reproduce the total uncertainty given in Circular T. The statistical uncertainty ranged between 0.8×10160.8\times 10^{-16} and 1.5×10161.5\times 10^{-16}, and decreased to 4×10174\times 10^{-17} in the full campaign period.

Refer to caption
Figure 9: Absolute frequencies of the 171Yb clock transition using a link to TAI [66, 67, 68, 49, 19, 20] (red circles), the 171Yb/87Sr frequency ratio [69, 72, 70, 45, 28, 71] and the CIPM recommended frequency of 87Sr [8] (blue triangles), and local Cs fountain clocks [11, 15] (black diamonds). The shaded region indicates the CIPM recommended value of 171Yb and fCIPM(Yb)=518 295 836 590 863.6(26)f^{\mathrm{CIPM}}(\mathrm{Yb})=518\,295\,836\,590\,863.6(26) Hz [8]. NIST: National Institute of Standards and Technology, KRISS: Korea Research Institute of Standards and Science, INRiM: Istituto Nazionale di Ricerca Metrologica, PTB: Physikalisch-Technische Bundesanstalt, NICT: National Institute of Information and Communications Technology.

Table 3 summarizes the uncertainties for the absolute frequency measurement. The statistical uncertainty of the Yb clock was calculated using the stability of the Yb/Sr ratio measurement (see Fig. 5) with the duration of the operation. The gravitational red shift was estimated to be 22.9(6)×101622.9(6)\times 10^{-16} using the Yb height from the conventionally adopted geoid potential (62 636 856.0 m2/s2). The total statistical uncertainty in the full period was 1.8×10161.8\times 10^{-16} (dashed blue lines in Fig. 8), which was obtained by inflating the uncertainty of the weighted mean by the square root of the reduced chi-square χred2=1.3\sqrt{\chi^{2}_{\mathrm{red}}}=1.3. Finally, we obtained y(YbSI)=1.1(5.0)×1016y(\mathrm{Yb}-\mathrm{SI})=-1.1(5.0)\times 10^{-16}. The absolute frequency of the Yb clock transition was determined as 518 295 836 590 863.54(26)518\,295\,836\,590\,863.54(26) Hz. The total uncertainty was mostly limited by the systematic uncertainties of the Yb clock and the microwave synthesis. Our determined value agreed with the recommended frequency fCIPM(Yb)f^{\mathrm{CIPM}}(\mathrm{Yb}) and the results of previous measurements by many groups (see Fig. 9).

7 Discussions and conclusion

The uncertainty in the link between the Yb clock and the SI second was 2.4×10162.4\times 10^{-16}, which was calculated by using the quadratic sum of the uncertainties due to the dead times in both Yb - UTC(NMIJ) and TAI - SI, the satellite link, and the microwave synthesis. The reduction of the link uncertainty by the continuous operation also enables measurements of the frequency ratios at the 101610^{-16} level between remotely-located secondary representations of the second. For example, during the present half-year period, a Rb fountain clock (SYRTE FORb) [64] was running. The frequency difference data y(TAIRb)y(\mathrm{TAI}-\mathrm{Rb}) for 180 days were obtained from Circular T. With our data y(YbTAI)y(\mathrm{Yb}-\mathrm{TAI}), we determined the Yb/Rb ratio as 75 833.197 545 114 174(42)75\,833.197\,545\,114\,174(42) with a relative uncertainty of 5.5×10165.5\times 10^{-16}. This agreed with the ratio calculated from the recommended frequencies fCIPM(Yb)/fCPIM(Rb)=75 833.197 545 114 196(59)f^{\mathrm{CIPM}}(\mathrm{Yb})/f^{\mathrm{CPIM}}(\mathrm{Rb})=75\,833.197\,545\,114\,196(59) [8] and the value 75 833.197 545 114192(33)75\,833.197\,545\,114192(33) reported by another group [19]. As another example, a Sr optical lattice clock (NICT-Sr1 [29]) calibrated the TAI frequency for 20 days during the sixth campaign period. This enabled us to determine the Yb/Sr ratio as 1.207 507 039 343 337 82(75)1.207\,507\,039\,343\,337\,82(75) with a fractional uncertainty of 6.2×10166.2\times 10^{-16}. This was in agreement with the most accurate previous value 1.207 507 039 343 337 749(55)1.207\,507\,039\,343\,337\,749(55) [45] and the other values with uncertainties at the 101610^{-16} level [28, 70, 71, 20].

The expectation value of the uptime of the Yb clock was calculated under the condition where the Yb clock is fully unattended during a night interval of 8 hours, and where long downtimes resulting from rare events such as facility maintenance are not included. The number of interrupting events that required manual recovery was 580580 during the full campaign period. This was obtained by counting the number of time gaps of more than 1 minute in the frequency data (Fig. 2). Since the Yb clock was operated with a 80.3%80.3\% uptime for 185 days, the mean number of interruptions per hour was 580/(0.803×185×24hours)=0.16580/(0.803\times 185\times 24\,\mathrm{hours})=0.16/hour. To estimate the probability of the occurrence of nn interrupting events in an interval of TT hours, we employed a Possison distribution,

P(n,T)=(0.16T)ne0.16Tn!.P(n,T)=\frac{(0.16T)^{n}e^{-0.16T}}{n!}. (3)

To calculate the expected uptime, the typical time for manually restarting the operation is needed. Figure 10 shows a histogram of the recovery time, which was obtained from time gaps in the frequency data (Fig. 2). Excluding long gaps exceeding half a day, the mean recovery time was 0.720.72 hours. The expected dead time during a 16-hour period of human work was then calculated as n=0nP(n,16)×0.72hours=1.8\sum_{n=0}^{\infty}nP(n,16)\times 0.72\,{\mathrm{hours}}=1.8 hours. During an 8-hour night period, the operation is stopped after one interrupting event. In this case, the expected dead time is calculated as 08d(1P(0,T))dT(8T)𝑑T=3.5\int_{0}^{8}\frac{d(1-P(0,T))}{dT}(8-T)dT=3.5 hours. The uptime per day was therefore expected to be 78%78\%.

Refer to caption
Figure 10: Histogram of the recovery time of more than 1 minute. The inset shows recovery times within half a day with a different bin size, from which the mean recovery time was obtained as 0.72 hours.

To consider the optimum conditions for the operation, we carried out a simulation of the dead time uncertainty as a function of the clock uptime (see Figure 11). In this simulation, the dead time was homogeneously distributed for a 30-day period, except for a large time gap of Tgap=15T_{\mathrm{gap}}=1-5 days in the middle of the period. Without the large gap (indicated as “None” in Fig. 11), the dead time uncertainty falls below 101610^{-16} for uptimes of 80%\geq 80\%. When the clock operation is stopped for Tgap=15T_{\mathrm{gap}}=1-5 days, the uncertainty no longer reaches below 101610^{-16} even though the uptime for the overall period is 80%\geq 80\%. This indicates that the uncertainties calculated with the actual duration of the clock operation are mostly determined by the instabilities of UTC(NMIJ) during the large gaps. When we take into account the fact that the uncertainty of the absolute frequency measurement is limited by the systematic uncertainties of the Yb clock and the microwave synthesis (the combined uncertainty is 4.6×10164.6\times 10^{-16}), it may be enough to reduce the dead time uncertainty to 2×1016\sim 2\times 10^{-16}. In this case, the optimum condition is to operate the clock every day with a moderate uptime of 50%\sim 50\%. This is well below our expectation value of the uptime (78%78\%, see above). It should be noted that daily repetition of the startup and shutdown of the clock increases human effort, and thus it is better in practice to let the clock run continuously.

Refer to caption
Figure 11: Simulated dead time uncertainty as a function of the clock uptime for 30 days. As shown in the top figure, the dead time is homogeneously distributed over the 30-day period, except for a large gap time of Tgap=15T_{\mathrm{gap}}=1-5 days in the middle of the period.

Previous studies have demonstrated an improvement in local timescales by steering flywheel oscillators to intermittently operated lattice clocks [13, 29, 30, 31]. In these methods, the stabilities of the timescales are limited by the fluctuations of the flywheels during the dead time of the clocks, and thus can be further improved by continuous operation. An analysis for our local timescale will be presented in a future publication.

The performance at the present level can make a significant contribution to the search for new physics. For example, local position invariance has been tested by looking for an annual variation of the gravitational red shift of atomic clocks [22, 12, 19]. As demonstrated in previous experiments that utilized continuously running fountain clocks and hydrogen masers [73], the continuous operation contributes to reduce the statistical uncertainty in a relatively short term, resulting in a stringent limit on violation of local position invariance. In our case, this experiment can be carried out by comparing the Yb clock with a local Cs fountain clock [74] or a hydrogen maser. Another example is that transient variations of the fine structure constant resulting from the passage of dark matter may be observed in a global network of optical clocks [24]. The detection of the transient effects relies on correlations in the frequency data of worldwide clocks running simultaneously [75]. To obtain a large amount of temporally overlapped data from laboratories worldwide, it is highly desirable that the participating clocks are capable of running continuously for a long period. The continuous operation also makes it possible to search for dark matter interactions on long timescales that could not be studied in previous experiments [75, 76].

In conclusion, we have demonstrated the nearly continuous operation of an Yb optical lattice clock with an uptime of 80.3%80.3\% for half a year. This enabled us to link the Yb clock to the SI second with an uncertainty of 2.4×10162.4\times 10^{-16}, and to determine the absolute frequency of the Yb clock transition with a total uncertainty of 5.0×10165.0\times 10^{-16}. The present performance preliminary demonstrated that the robustness of optical lattice clocks can reach a level comparable to that of Cs fountain clocks. This work constitutes an important step towards the redefinition of the SI second.

Acknowledgments

We thank H. Katori, M. Takamoto, and H. Imai for providing information on their vacuum systems, and S. Okubo for technical assistance. We are indebted to national metrology institutes for their efforts in operating the primary and secondary frequency standards and making the data available in Circular T. This work was supported by Japan Society for the Promotion of Science (JSPS) KAKENHI Grant Number 17H01151, and 17K14367, and JST-Mirai Program Grant Number JPMJMI18A1, Japan.

Reference

References

  • [1] Bloom B J, Nicholson T L, Williams J R, Campbell S L, Bishof M, Zhang X, Zhang W, Bromley S L, and Ye J 2014 An optical lattice clock with accuracy and stability at the 101810^{-18} level NatureNature 𝟓𝟎𝟔\bf{506} 71-75
  • [2] Ushijima I, Takamoto M, Das M, Ohkubo T, and Katori H 2015 Cryogenic optical lattice clocks Nat.Nat. Photon.Photon. 𝟔\bf{6} 185-189
  • [3] Nicholson T L, Campbell S L, Hutson R B, Marti G E, Bloom B J, McNally R L, Zhang W, Barrett M D, Safronova M S, Strouse G F, Tew W L, and Ye J 2015 Systematic evaluation of an atomic clock at 2×10182\times 10^{-18} total uncertainty Nat.Nat. Commun.Commun. 𝟔\bf{6} 6896
  • [4] McGrew W F, Zhang X, Fasano R J, Schäffer S A, Beloy K, Nicolodi D, Brown R C, Hinkley N, Milani G, Schioppo M, Yoon T H, and Ludlow A D 2018 Atomic clock performance enabling geodesy below the centimetre level NatureNature 𝟓𝟔𝟒\bf{564} 87-90
  • [5] Bothwell T, Kedar D, Oelker E, Robinson J M, Bromley S L, Tew W L, Ye J, and Kennedy C J 2019 JILA SrI optical lattice clock with uncertainty of 2.0×10182.0\times 10^{-18} MetrologiaMetrologia 𝟓𝟔\bf{56} 066004.
  • [6] Gill P 2011 When should we change the definition of the second? Philos.Philos. Trans.Trans. Roy.Roy. Soc.Soc. A 𝟑𝟔𝟗\bf{369} 4109-4130
  • [7] Hong F-L 2016 Optical frequency standards for time and length applications Meas.Meas. Sci.Sci. Technol.Technol. 𝟐𝟖\bf{28} 012002
  • [8] Riehle F, Gill P, Arias F, and Robertsson L. 2018 The CIPM list of recommended frequency standard values: Guidelines and procedures MetrologiaMetrologia 𝟓𝟓\bf{55} 188-200
  • [9] Lodewyck J 2019 On a definition of the SI second with a set of optical clock transitions MetrologiaMetrologia 𝟓𝟔\bf{56} 055009
  • [10] Campbell G K, Ludlow A D, Blatt S, Thomsen J W, Martin M J, de Miranda M H G, Zelevinsky T, Boyd M M, Ye J, Diddams S A, Heavner T P, Parker T E, and Jefferts S R 2008 The absolute frequency of the 87Sr optical clock transition MetrologiaMetrologia 𝟒𝟓\bf{45} 539-548
  • [11] Lemke N D, Ludlow A D, Barber Z W, Fortier T M, Diddams S A, Jiang Y, Jefferts S R, Heavner T P, Parker T E, and Oates C W 2009 Spin-1/2 optical lattice clock Phys.Phys. Rev.Rev. Lett.Lett. 𝟏𝟎𝟑\bf{103} 063001
  • [12] Le Targat R, Lorini L, Le Coq Y, Zawada M, Gueńa J, Abgrall M, Gurov M, Rosenbusch P, Rovera D G, Nagórny B, Gartman R, Westergaard P G, Tobar M E, Lours M, Santarelli G, Clairon A, Bize S, Laurent P, Lemonde P, and Lodewyck J 2013 Experimental realization of an optical second with strontium lattice clocks Nat.Nat. Commun.Commun. 𝟒\bf{4} 2109
  • [13] Grebing C, Al-Masoudi A, Dörsher S, Häfner S, Gerginov V, Weyers S, Lipphardt B, Riehle F, Sterr U, and Lisdat C 2016 Realization of a timescale with an accurate optical lattice clock OpticaOptica 𝟑\bf{3} 563-569
  • [14] Lodewyck J, Bilicki S, Bookjans E, Robyr J-L, Shi C, Vallet G, Le Targat R, Nicolodi D, Coq Y, Guéna J, Abgrall M, Rosenbusch P, and Bize S 2016 Optical to microwave clock frequency ratios with a nearly continuous strontium optical lattice clock MetrologiaMetrologia 𝟓𝟑\bf{53} 1123-1130
  • [15] Pizzocaro M, Thoumany P, Rauf B, Bregolin F, Milani G, Clivati C, Costanzo G A, Levi F, and Calonico D 2017 Absolute frequency measurement of the 1S30{}_{0}-^{3}P0 transition of 171Yb MetrologiaMetrologia 𝟓𝟒\bf{54} 102-112
  • [16] Panfilo G and Parker T E 2010 A theoretical and experimental analysis of frequency transfer uncertainty, including frequency transfer into TAI MetrologiaMetrologia 𝟒𝟕\bf{47} 552-560
  • [17] Tanabe T, Akamatsu D, Kobayashi T, Takamizawa A, Yanagimachi S, Ikegami T, Suzuyama T, Inaba H, Okubo S, Yasuda M, Hong F-L, Onae A, and Hosaka K 2015 Improved frequency measurement of the 1S30{}_{0}-^{3}P0 clock transition in 87Sr using a Cs fountain clock as a transfer oscillator JJ. PhysPhys. SocSoc. JpnJpn. 𝟖𝟒\bf{84} 115002
  • [18] Hachisu H, Petit G, Nakagawa F, Hanado Y, and Ido T 2017 SI-traceable measurement of an optical frequency at the low 101610^{-16} level without a local primary standard OptOpt. ExpressExpress 𝟐𝟓\bf{25} 8511-8523
  • [19] McGrew W F, Zhang X, Leopardi H, Fasano R J, Nicolodi D, Beloy K, Yao J, Sherman J A, Schäffer S A, Savory J, Brown R C, Römisch S, Oates C W, Parker T E, Fortier T M, and Ludlow A D 2019 Towards the optical second: verifying optical clocks at the SI limit OpticaOptica 𝟔\bf{6} 448-454
  • [20] Pizzocaro M, Bregolin F, Barbieri P, Rauf B, Levi F, and Calonico D 2019 Absolute frequency measurement of the 1S30{}_{0}-^{3}P0 transition of 171Yb with a link to International Atomic Time MetrologiaMetrologia
  • [21] Hill I R, Hobson R, Bowden W, Bridge E M, Donnellan S, Curtis E A, and Gill P 2016 A low maintenance Sr optical lattice clock JJ. PhysPhys. ConfConf. SerSer. 𝟕𝟐𝟑\bf{723} 012019
  • [22] Blatt S, Ludlow A D, Campbell G K, Thomsen J W, Zelevinsky T, Boyd M M, Ye J, Baillard X, Fouché M, Le Targat R, Brusch A, Lemonde P, Takamoto M, Hong F-L, Katori H, Flambaum V V 2008 New Limits on Coupling of Fundamental Constants to Gravity Using 87Sr Optical Lattice Clocks Phys.Phys. Rev.Rev. Lett.Lett. 𝟏𝟎𝟎\bf{100} 140801
  • [23] Safronova M S, Budker D, DeMille D, Jackson Kimball D F, Derevianko A, and Clark C W 2018 Search for new physics with atoms and molecules Rev.Rev. Mod.Mod. Phys.Phys. 𝟗𝟎\bf{90} 025008
  • [24] Derevianko A and Pospelov M 2014 Hunting for topological dark matter with atomic clocks Nat.Nat. Phys.Phys. 𝟏𝟎\bf{10} 933-936
  • [25] Wcisło P, Morzyński P, Bober M, Cygan A, Lisak D, Ciuryło R, and Zawada M 2016 Experimental constraint on dark matter detection with optical atomic clocks Nat.Nat. Astron.Astron. 𝟏\bf{1} 0009
  • [26] Takamoto M, Ushijima I, Ohmae N, Yahagi T, Kokado K, Shinkai H, and Katori H 2020 Test of General Relativity by a Pair of Transportable Optical Lattice Clocks NatNat. PhotonPhoton.
  • [27] Takano T, Takamoto M, Ushijima I, Ohmae N, Akatsuka T, Yamaguchi A, Kuroishi, Munekane H, Miyahara B, and Katori H 2016 Geopotential measurements with synchronously linked optical lattice clocks Nat.Nat. Photon.Photon. 𝟏𝟎\bf{10} 662-666
  • [28] Grotti J etet. alal. 2018 Geodesy and metrology with a transportable optical clock Nat.Nat. Phys.Phys. 𝟏𝟒\bf{14} 437-441
  • [29] Hachisu H, Nakagawa F, Hanado Y, and Ido T 2018 Months-long real-time generation of a time scale based on an optical clock Sci.Sci. Rep.Rep. 𝟖\bf{8} 4243
  • [30] Milner W R, Robinson J M, Kennedy C J, Bothwell T, Kedar D, Matei D G, Legero T, Sterr U, Riehle F, Leopardi H, Fortier T M, Sherman J A, Levine J, Yao J, Ye J, and Melker E 2019 Demonstration of a Timescale Based on a Stable Optical Carrier Phys.Phys. Rev.Rev. Lett.Lett. 𝟏𝟐𝟑\bf{123} 173201
  • [31] Yao J, Sherman J A, Fortier T, Leopardi H, Parker T, McGrew W, Zhang X, Nicolodi D, Fasano R, Schäffer S, Beloy K, Savory J, Romisch S, Oates C, Diddams S, Ludlow A, and Levine J 2019 Optical-Clock-Based Time Scale Phys.Phys. Rev.Rev. AppliedApplied 𝟏𝟐\bf{12} 044069
  • [32] Inaba H, Daimon Y, Hong F-L, Onae A, Minoshima K, Schibli T R, Matsumoto H, Hirano M, Okuno T, Onishi M, and Nakazawa M 2006 Long-term measurement of optical frequencies using a simple, robust and low-noise fiber based frequency comb Opt.Opt. ExpressExpress 𝟏𝟒\bf{14} 5223-5231
  • [33] Hisai Y, Akamatsu D, Kobayashi T, Okubo S, Inaba H, Hosaka K, Yasuda M, and Hong F-L 2019 Development of 8-branch Er:fiber frequency comb for Sr and Yb optical lattice clocks Opt.Opt. ExpressExpress 𝟐𝟕\bf{27} 6404-6414
  • [34] Kobayashi T, Akamatsu D, Hosaka K, and Yasuda M 2019 A relocking scheme for optical phase locking using a digital circuit with an electrical delay line Rev.Rev. Sci.Sci. Instrum.Instrum. 𝟗𝟎\bf{90} 103002
  • [35] Kobayashi T, Akamatsu D, Hisai Y, Tanabe T, Inaba H, Suzuyama T, Hong F-L, Hosaka K, and Yasuda M 2018 Uncertainty Evaluation of an 171Yb Optical Lattice Clock at NMIJ IEEEIEEE Trans.Trans. Ultrason.Ultrason. Ferroelectr.Ferroelectr. Freq.Freq. ControlControl 𝟔𝟓\bf{65} 2449-2458
  • [36] Katori H, Takamoto M, Pal’chikov V G, and Ovsiannikov V D 2003 Ultrastable optical clock with neutral atoms in an engineered light shift trap Phys.Phys. Rev.Rev. Lett.Lett. 𝟗𝟏\bf{91} 173005
  • [37] Kobayashi T, Akamatsu D, Hosaka K, Inaba H, Okubo S, Tanabe T, Yasuda M, Onae A, and Hong F-L 2016 Absolute frequency measurements and hyperfine structures of the molecular iodine transitions at 578 nm J.J. Opt.Opt. Soc.Soc. Am.Am. BB 𝟑𝟑\bf{33} 725-734
  • [38] Saxberg B, Plotkin-Swing B, and Gupta S Active stabilization of a diode laser injection lock 2016 Rev.Rev. Sci.Sci. Instrum.Instrum. 𝟖𝟕\bf{87} 063109
  • [39] Kobayashi T, Akamatsu D, Nishida Y, Tatabe T, Yasuda M, Hong F-L, and Hosaka K 2016 Second harmonic generation at 399 nm resonant on the S011P0{}^{1}S_{0}-^{1}P_{0} transition of ytterbium using a periodically poled LiNbO3 waveguide Opt.Opt. ExpressExpress 𝟐𝟒\bf{24} 12142-12150
  • [40] Iwakuni K, Inaba H, Nakajima Y, Kobayashi T, Hosaka K, Onae A, and Hong F-L 2012 Narrow linewidth comb realized with a mode-locked fiber laser using an intra-cavity waveguide electro-optic modulator for high-speed control Opt.Opt. ExpressExpress 𝟐𝟎\bf{20} 13769-13776
  • [41] Inaba H, Hosaka K, Yasuda M, Nakajima Y, Iwakuni K, Akamatsu D, Okubo S, Kohno T, Onae A, and Hong F-L 2013 Spectroscopy of 171Yb in an optical lattice based on laser linewidth transfer using a narrow linewidth frequency comb Opt.Opt. ExpressExpress 𝟐𝟏\bf{21} 7891-7896
  • [42] Hisai Y, Ikea K, Sakagami H, Horikiri T, Kobayashi T, Yoshii K, and Hong F-L 2018 Evaluation of laser frequency offset locking using an electrical delay line Appl.Appl. Opt.Opt. 𝟓𝟕\bf{57} 5628-5634
  • [43] Akamatsu D, Inaba H, Hosaka K, Yasuda M, Onae A, Suzuyama T, Amemiya M, and Hong F-L 2013 Spectroscopy and frequency measurement of the 87Sr clock transition by laser linewidth transfer using an optical frequency comb Appl.Appl. Phys.Phys. ExpressExpress 𝟕\bf{7} 012401
  • [44] Katori H, Ovsiannikov V D, Marmo S I, and Palchikov V G 2015 Strategies for reducing the light shift in atomic clocks Phys.Phys. Rev.Rev. AA 𝟗𝟏\bf{91} 052503
  • [45] Nemitz N, Ohkubo T, Takamoto M, Ushijima I, Das M, Ohmae N, and Katori H 2016 Frequency ratio of Yb and Sr clocks with 5×10175\times 10^{-17} uncertainty at 150 seconds averaging time Nat.Nat. Photon.Photon. 𝟏𝟎\bf{10} 258-261
  • [46] Nemitz N, Jørgensen A A, Yanagimoto R, Bregolin F, and Katori H 2019 Modeling light shifts in optical lattice clocks Phys.Phys. Rev.Rev. AA 𝟗𝟗\bf{99} 033424
  • [47] Brown R C, Phillips N B, Beloy K, McGrew W F, Schioppo M, Fasano R J, Milani G, Zhang X, Hinkley N, Leopardi H, Yoon T H, Nicolodi D, Fortier T M, and Ludlow A D 2017 Hyperpolarizability and Operational Magic Wavelength in an Optical Lattice Clock Phys.Phys. Rev.Rev. Lett.Lett. 𝟏𝟏𝟗\bf{119} 253001
  • [48] Blatt S, Thomsen J W, Campbell G K, Ludlow A D, Swallows M D, Martin M J, Boyd M M, and Ye J 2009 Rabi spectroscopy and excitation inhomogeneity in a one-dimensional optical lattice clock Phys.Phys. Rev.Rev. AA 𝟖𝟎\bf{80} 052703
  • [49] Kim H, Heo M-S, Lee W-K, Park C Y, Hong H-G, Hwang S-W, and Yu D-H 2017 Improved absolute frequency measurement of the 171Yb optical lattice clock at KRISS relative to the SI second Jpn.Jpn. J.J. Appl.Appl. Phys.Phys. 𝟓𝟔\bf{56} 050302
  • [50] Middelmann T, Lisdat C, Falke S, Vellore Winfred J S R, Riehle F, and Sterr U 2011 Tackling the blackbody shift in a strontium optical lattice clock IEEEIEEE Trans.Trans. Instum.Instum. Meas.Meas. 𝟔𝟎\bf{60} 2550
  • [51] Sherman J A, Lemke N D, Hinkley N, Pizzocaro M, Fox R W, Ludlow A D, and Oates C W 2012 High-Accuracy Measurement of Atomic Polarizability in an Optical Lattice Clock Phys.Phys. Rev.Rev. Lett.Lett. 𝟏𝟎𝟖\bf{108} 153002
  • [52] Beloy K, Hinkley N, Phillips N B, Sherman J A, Shiopppo M, Lehman J, Feldman A, Hanssen L M, Oates C W, Ludlow A D 2014 Atomic Clock with 1×10181\times 10^{-18} Room-Temperature Blackbody Stark Uncertainty Phys.Phys. Rev.Rev. Lett.Lett. 𝟏𝟏𝟑\bf{113} 260801
  • [53] Lemke N D, von Stecher J, Sherman J A, Rey A M, Oates C W, and Ludlow A D 2011 pp-Wave Cold Collisions in an Optical Lattice Clock Phys.Phys. Rev.Rev. Lett.Lett. 𝟏𝟎𝟕\bf{107} 103902
  • [54] Ludlow A D, Lemke N D, Sherman J A, Oates C W, Quéméner G, von Stecher J, and Rey A M 2011 Cold-collision-shift cancellation and inelastic scattering in a Yb optical lattice clock Phys.Phys. Rev.Rev. AA 𝟖𝟒\bf{84} 052724
  • [55] Lee S, Park C Y, Lee W-K, and Yu D-H 2016 Cancellation of collisional frequency shifts in optical lattice clocks with Rabi spectroscopy NewNew J.J. Phys.Phys. 𝟏𝟖\bf{18} 033030
  • [56] Nakajima Y, Inaba H, Hosaka K, Minoshima K, Onae A, Yasuda M, Kohno T, Kawato S, Kobayashi T, Katsuyama T, and Hong F-L 2010 A multi-branch, fiber-based frequency comb with millihertz-level relative linewidths using an intra-cavity electro-optic modulator Opt.Opt. ExpressExpress 𝟏𝟖\bf{18} 1667-1676
  • [57] https://www.bipm.org/en/bipm/tai/.
  • [58] Yu D-H, Weisse M, and Parker T E 2007 Uncertainty of a frequency comparison with distributed dead time and measurement interval offset MetrologiaMetrologia 𝟒𝟒\bf{44} 91-96
  • [59] Bauch A 2005 The PTB primary clocks CS1 and CS2 MetrologiaMetrologia 𝟒𝟐\bf{42} S43-S54
  • [60] Guéna J, Abgrall M, Rovera D, Laurent P, Chupin B, Lours M, Santarelli G, Rosenbusch P, Tobar M E, Li R, Gibble K, Clairon A, and Bize S 2012 Progress in atomic fountains at LNE-SYRTE IEEEIEEE Trans.Trans. Ultrason.Ultrason. Ferroelectr.Ferroelectr. Freq.Freq. ControlControl 𝟓𝟗\bf{59} 391-409
  • [61] Weyers S, Gerginov V, Kazda M, Rahm J, Lipphardt B, Dobrev G, and Gibble K 2018 Advances in the accuracy, stability, and reliability of the PTB primary fountain clocks MetrologiaMetrologia 𝟓𝟓\bf{55} 789-805
  • [62] Blinov I Y, Boiko A I, Domnin Y S, Kostromin V P, Kupalova O V, and Kupalov D S 2017 Budget of Uncertainties in the Cesium Frequency Frame of Fountain Type MeasurementMeasurement TechniquesTechniques 𝟔𝟎\bf{60} 30-36
  • [63] Fang F, Li M, Lin P, Chen W, Liu N, Lin Y, Wang P, Liu K, Suo R, and Li T 2015 NIM5 Cs fountain clock and its evaluation MetrologiaMetrologia 𝟓𝟐\bf{52} 454-468
  • [64] Guéna J, Abgrall M, Clairon A, and Bize S 2014 Contributing to TAI with a secondary representation of the SI second MetrologiaMetrologia 𝟓𝟏\bf{51} 108-120
  • [65] Hachisu H, Petit G, and Ido T 2016 Absolute frequency measurement with uncertainty below 1×10151\times 10^{-15} using International Atomic Time Appl.Appl. Phys.Phys. BB 𝟏𝟐𝟑\bf{123} 34
  • [66] Kohno T, Yasuda M, Hosaka K, Inaba H, Nakajima Y, and Hong F-L 2009 One-Dimensional Optical Lattice Clock with a Fermionic 171Yb Isotope Appl.Appl. Phys.Phys. ExpressExpress 𝟐\bf{2} 072501
  • [67] Yasuda M, Inaba H, Kohno T, Tanabe T, Nakajima Y, Hosaka K, Akamatsu D, Onae A, Suzuyama T, Amemiya M, and Hong F-L 2012 Improved Absolute Frequency Measurement of the 171Yb Optical Lattice Clock towards a Candidate for the Redefinition of the Second Appl.Appl. Phys.Phys. ExpressExpress 𝟓\bf{5} 102401
  • [68] Park C Y, Yu D-J, Lee W-K, Park S E, Kim E B, Lee S K, Cho J W, Yoon T H, Mun J, Park S J, Kwon T Y, and Lee S-B 2013 Absolute frequency measurement of 1S(F=1/2)03{}_{0}(F=1/2)-^{3}P(F=1/2)0{}_{0}(F=1/2) transition of 171Yb atoms in a one-dimensional optical lattice at KRISS MetrologiaMetrologia 𝟓𝟎\bf{50} 119-128
  • [69] Akamatsu D, Yasuda M, Inaba H, Hosaka K, Tanabe T, Onae A, and Hong F-L 2014 Frequency ratio measurement of 171Yb and 87Sr optical lattice clocks Opt.Opt. ExpressExpress 𝟐𝟐\bf{22} 7898-7905
  • [70] Takamoto M, Ushijima I, Das M, Nemitz N, Ohkubo T, Yamanaka K, Ohmae N, Takano T, Akatsuka T, Yamaguchi A, and Katori H 2015 Frequency ratios of Sr, Yb, and Hg based optical lattice clocks and their applications C.C. R.R. Phys.Phys. 𝟏𝟔\bf{16} 489-498
  • [71] Fujieda M, Yang S-H, Gotoh T, Hwang S-W, Hachisu H, Kim H, Lee Y K, Tabuchi R, Ido T, Lee W-K, Heo M-S, Park C Y, Yu D-H, and Petit G 2018 Advanced satellite-based frequency transfer at the 101610^{-16} level IEEEIEEE Trans.Trans. Ultrason.Ultrason. Ferroelectr.Ferroelectr. Freq.Freq. ControlControl 𝟔𝟓\bf{65} 973-978
  • [72] Akamatsu D, Kobayashi T, Hisai Y, Tanabe T, Hosaka K, Yasuda M, and Hong F-L 2018 Dual-mode operation of an optical lattice clock using strontium and ytterbium atoms IEEEIEEE Trans.Trans. Ultrason.Ultrason. Ferroelectr.Ferroelectr. Freq.Freq. ControlControl 𝟔𝟓\bf{65} 1069-1075
  • [73] Peil S, Crane S, Hanssen J L, Swanson T B, and Ekstrom C R 2013 Tests of local position invariance using continuously running atomic clocks Phys.Phys. Rev.Rev. AA 𝟖𝟕\bf{87} 010102(R)
  • [74] Takamizawa A, Yanagimachi S, Tanabe T, Hagimoto K, Hirano I, Watabe K, Ikegami T, and Hartnett J G 2015 Preliminary Evaluation of the Cesium Fountain Primary Frequency Standard NMIJ-F2 IEEEIEEE Trans.Trans. Instrum.Instrum. Meas.Meas. 𝟔𝟒\bf{64} 2504-2512
  • [75] Wcisło P etet al.al. 2018 New bounds on dark matter coupling from a global network of optical atomic clocks Sci.Sci. Adv.Adv. 𝟒\bf{4}, eaau4869
  • [76] Roberts B M etet al.al. 2019 Search for transient variations of the fine structure constant and dark matter using fiber-linked optical atomic clocks arXiv:1907.02661